Print display data
Field: Weight Mode Stab CR LF
Length: 9 1 5 1 1 1
When "numeric only"data is
selected for output in the
RS232 menu, the Mode field
is not output.
Command
Command
Command
Command
Command
Character
Character
Character
Character
Character
Description
Description
Description
Description
Description
C
C
C
C
C
Begin span calibration
xD
xD
xD
xD
xD
Set 1 second print delay (set x = 0 for OFF, or x = 1 for ON)
F
F
F
F
F
Print current function.
x
x
x
x
x
IIIII
Set Averaging Filter Level to “x”,
where x = 0 to 3 (see table).
If LFT, level 0 to 1.
L
L
L
L
L
Begin linearity calibration
xM
xM
xM
xM
xM
Places balance in mode “x”,
where x = 1 to 17 (see table).
If unit or mode is not already
enabled, command will be ignored.
RS232 COMMAND TABLE (Cont.)
0
=
minimum level
1
=
2
=
3
=
maximum level
1
=
milligrams
2
=
grams
3
=
kilo grams
4
=
dwt
5
=
Carats
6
=
Ounces
7
=
Ounces troy
8
=
Grains
9
=
Taels Hong Kong
10
=
Taels Singapore
11
=
Taels Taiwan
12
=
Mommes
13
=
Decimal Pounds
14
=
(Not used)
15
=
Newton's
16
=
tical
17
=
Custom Units
Same as ?
command
Displayed weight sent right justified
w/lead zero blanking.
Nine characters include:
decimal point (1)
weight (7 max)
polarity (1): blank if positive
"-" if negative
xSL
xSL
xSL
xSL
xSL
Set stable data only printing (set x = 0 to 3). If LFT 0 or 1.
T
T
T
T
T
Same effect as pressing O/T button.
V
V
V
V
V
Print EPROM version
Esc V
Esc V
Esc V
Esc V
Esc V
Print balance ID (13 characters).
xZ
xZ
xZ
xZ
xZ
Set Auto Zero to “x”,where x = 0 to 3 ). 0=Off, 1=0.5d, 2=1d, 3=5d.If LFT,
programs Auto zero level from 0 to 1.
x%
x%
x%
x%
x%
Set % reference function. Uses x (Real Number) as current Percent Reference.
Reference weigh must be entered in grams.
x#
x#
x#
x#
x#
Set PC reference function Uses x (Real Number) as current Percent Reference.
Reference weigh must be entered in grams.
NOTE: Availability of shaded units subject to local regulations.