3.5 Percent Weighing

Percent Weighing is

 enabled only when Percent is turned ON in the Mode submenu under Setup. Percent

weighing permits you to place a reference load on the balance, then view other loads as a percentage of the
reference. The load you place on the pan as a reference may be displayed as any percentage you select from 5 %
to 100 % (in 1 % increments). One hundred percent does not necessarily have to represent the reference load.
Subsequent loads, displayed as a percentage of the reference are  limited only by the capacity of the balance.
The default setting is Reference 100 %. Refer to paragraph 4.8 to enable percent weighing.

Stable 

%    

Mode

Use          keys to change,

press Enter

Change 

Mode

Use          keys to change,

press Enter

Change 

%    

Procedure

• Press the Mode button.

• Press 

 or 

 button until PERCENT is dis-

played.

• Press Enter button, PUT>PAN 100 % is displayed.

If a container is used, the balance can be tared at
this point. The % display momentarily blanks while
the balance is taring out.

• Put the reference load on the pan.

• Press 

 or 

 button and select reference

weight percentage (Percent Range 5 to 100). Hold

button down for fast change.

• Press Enter button to save setting, WORKING is

displayed ... calculating reference weight.

Balance displays reference weight for five seconds in
selected measuring unit, then displays the percentage.

• Remove the reference weight from the pan and

replace it with another load. The second load is dis-
played as a percentage of the reference.

NOTE: The PERCENT display (number of digits) is a
function of the accuracy of the balance and the size of
the reference weight. The display examples were with a
300 g mass used with an 8.1 kilogram balance.

Stable 

   g

Содержание EXPLORER

Страница 1: ...INSTRUCTION MANUAL Balances O T O T OHAUS Stable g Print Setup Units Mode Enter O T O T OHAUS Stable g Print Setup Units Mode Enter...

Страница 2: ......

Страница 3: ...ed input current or 16A Limits Section 2 Limits for harmonic current emissions Limits section 3 Limitation of voltage fluctuations and flicker in low voltage supply systems Ohaus Corporation Pine Broo...

Страница 4: ...LE MINISTRE DES Unauthorized changes or modifications to this equipment are not permitted EC DECLARATION OF CONFORMITY Explorer E10640 E00640 E10645 E00645 E11140 E01140 E11145 E01145 E12140 E02140 E...

Страница 5: ...of the Instrument 6 3 2 Turning On the Balance 7 3 3 Calibration 7 3 3 1 Internal Calibration InCALTM 9 3 3 2 Calibration Message 10 3 3 3 Calibration Adjust 11 3 3 4 Span Calibration 12 3 3 5 User Ca...

Страница 6: ...Units 30 4 10 Global 30 4 11 Custom Unit 32 4 12 Menu Lock Out Protection 34 5 CARE AND MAINTENANCE 35 5 1 Troubleshooting 35 5 2 RS232 Interface 36 5 3 Error Codes List 39 5 4 Information Messages 3...

Страница 7: ...ighing units 4 Setup button Selects various submenus calibration date time readout GLP data GLP set print RS232 LFT function units global custom 5 O T button When pressed performs tare function and a...

Страница 8: ...w what area of the balance menu is being addressed 10 Units Is displayed when the Units button is pressed Allows the user to know what area of the balance menu is being addressed 11 Setup Is displayed...

Страница 9: ...ds Remote displays upgraded displays which can be table wall or tower mounted are available as accessories It offers a high level of operating convenience and useful functions to make accurate measure...

Страница 10: ...zontal positioning and stable installation are prerequisites for repeatable results To compensate small irregularites or inclinations at the location the instrument can be leveled For exact horizontal...

Страница 11: ...is in place air currents are deflected up over the pan Make sure the wind shield is firmly snapped into place NOTE 4100 g 6100 g and 8100 g balances with internal calibration are equipped with a 6 Pa...

Страница 12: ...ings which are bolded The setup submenus shown below are arranged in the order as displayed in the balance CALIBRATION InCALTM Calibration No Lock Span Calibration User Calibration Linearity Calibrati...

Страница 13: ...lass I balances require at least 2 hours The internal circuits of the balance are powered whenever it is plugged into a power source Calibration Refer to paragraph 3 3 and calibrate the balance before...

Страница 14: ...in previously stored calibration data Calibration should be performed as necessary to ensure accurate weighing Masses required to perform the procedures are listed in the adjacent table NOTE Any of th...

Страница 15: ...r Change Procedure Press the Setup button CAL is displayed Press Enter button CAL TYPE is displayed Press Enter button CAL TYPE InCAL is displayed Press Enter button INCAL is displayed NOTE If a weigh...

Страница 16: ...e display NOTE When the balance is in a legal for trade mode this option is locked ON In this condition the balance will display the CAL NOW message when recalibration is required Press Enter button S...

Страница 17: ...icates LOCK Press Enter button LOCK ON or LOCK OFF is dis played Press or button and select either LOCK ON or LOCK OFF LOCK ON is normally used in legal for trade applications When set ON and legal fo...

Страница 18: ...hange Setup press Enter Change g Setup Change Procedure Press the Setup button CAL is displayed Press Enter button CAL TYPE is displayed Press Enter button CAL TYPE InCAL is displayed Press button to...

Страница 19: ...e to select USER calibration CAL TYPE USer is displayed Press Enter button the display indicates the last calibration mass value which was entered with the first digit flashing Sample illustrates 2000...

Страница 20: ...an Press Enter button WORKING is displayed After a few seconds CAL SET is displayed the display then returns to WEIGH mode Linearity calibration is completed Remove calibration mass from the pan Setup...

Страница 21: ...alibration Display changes to PUT WT 8000 g The displayed weight is the full capacity of the balance Place specified calibration mass on pan Press Enter button WORKING is displayed After a few seconds...

Страница 22: ...mpleted LIN CAL 12 01 97 1 00 00 PM Bal Id 1234 Cal 8000 00 g Old 7999 08 g Dif 0 02 g Wt Ref USER NO 2056853 PROJ NO 100012 Name END CAL TEST 12 01 97 1 00 00 PM Bal Id 1234 Cal 8000 00 g Act 8000 02...

Страница 23: ...measuring unit Press Enter button balance is now ready for weighing Place the object s or material to be weighed on the pan Example illustrates a 200 gram weight Wait for the stability indicator to a...

Страница 24: ...o change press Enter Change Procedure Press the Mode button Press or button until PERCENT is dis played Press Enter button PUT PAN 100 is displayed If a container is used the balance can be tared at t...

Страница 25: ...uired Press or button and select sample size Sample size is 5 to 1000 pieces Place sample size on the pan Press Enter button to continue display indicates WORKING Balance displays the reference weight...

Страница 26: ...sing any buttons When the balance displays READY simply place subject on pan Starting Animal Cycle Place animal container if used on pan Press to tare the container Place subject on pan The animal cyc...

Страница 27: ...l computer or printer requires that the communications parameters in the Setup menu Print options and communication parameters be set first Refer to page 25 Print menu settings and page 27 for RS232 c...

Страница 28: ...nter button SAVED displays momentarily and EXIT appears Press Enter button balance returns to a weighing mode Adjustments up to 60 seconds a month can be made to the balance internal clock Repeat the...

Страница 29: ...exceeded Factory default setting is shown in bold type OFF Turns Auto Zero OFF 5 d Sets threshold to 5 divisions 1 d Sets threshold to 1 division 3 d Sets threshold to 3 divisions Filter Filter compe...

Страница 30: ...e display and enter a 7 digit number for the project number Press Enter button to save setting Press or button to select next item or EXIT Press Enter button to continue Procedure To select any of the...

Страница 31: ...on to continue Press or button and select either menu setting or ON or OFF Press Enter button to save setting Press or button to select next item or EXIT Press Enter button to continue Auto Print Feat...

Страница 32: ...looked at but not changed in the menu When set off entries may be changed OFF is the default setting 12 01 97 12 01 37 AM 429 5 g 12 01 97 12 01 52 AM Bal Id 429 8 g 12 01 97 12 02 17 AM Bal Id USER N...

Страница 33: ...e number of data bits to 7 or 8 The default setting is 7 Stop Bits The number of stop bits can be set to 1 or 2 The default setting is 2 Lock Lock ON OFF can only be changed when the hardware Lock Swi...

Страница 34: ...balance to operate in a legal for trade application and works in conjunction with a Lock Switch Default setting is UNLOCKED See default table LFT and Lock Switch Menu lock Default Value Date Menu Unlo...

Страница 35: ...default Percent Percent weighing permits you to place a reference load on the balance then view other loads as a percentage of the reference Selection is made using the Mode button The default settin...

Страница 36: ...type of operation These functions are List and Reset The default settings are NO Global List is a convienent method of examining which parameters are set up in the balance The parameters do not show...

Страница 37: ...setting is OFF EXPLORER xxxxxx xxx Sr x xx OS x xx G x xx Time 12hr 3 19 51 PM Date m d y 12 01 97 Function Weigh ReadOut Menu Stb 2 d AZT 5 d Filter 1 GLP Menu Time Date On Bal Id On User No On Proj...

Страница 38: ...2 Moves decimal point 2 places to the left E 1 Moves decimal point 1 place to the left E0 Leaves decimal point in normal position E1 Moves decimal point 1 place to the right E2 Moves decimal point 2 p...

Страница 39: ...LSD is displayed There are 6 LSD least significant digit settings you can choose from see table Press Enter button LSD 1 is displayed Press or button and select LSD value either 1 2 5 10 or 100 Press...

Страница 40: ...l menus which have had LOCK turned ON The default setting for the Lock Switch is OFF Type Approved Legal for Trade Balance Sealing Certified balances have a securing sticker applied at the factory Whe...

Страница 41: ...rect weight reading Cannot display weight in desired unit Unable to store menu settings changes RS232 interface not working Random segments displayed or display locks up Unable to change settings Unst...

Страница 42: ...are shown in the adjacent illustration The balance will not output any data unless pin 5 CTS is held in an ON state 3 to 15 V dc Interfaces not utilizing the CTS handshake may tie pin 5 to pin 6 to de...

Страница 43: ...ms 3 kilo grams 4 dwt 5 Carats 6 Ounces 7 Ounces troy 8 Grains 9 Taels Hong Kong 10 Taels Singapore 11 Taels Taiwan 12 Mommes 13 Decimal Pounds 14 Not used 15 Newton s 16 tical 17 Custom Units Same as...

Страница 44: ...Program User ID String 1 8 characters AC AC AC AC AC Abort Calibration xUC xUC xUC xUC xUC User Weight Calibration IC IC IC IC IC Internal calibration InCALTM LE LE LE LE LE Show Last Error Code Resp...

Страница 45: ...calibration weight internal sensor indicated its weight on the pan CheckSum Errors 9 1 Bad factory checksum If error persists have the balance serviced 9 2 Bad factory checksum If error persists have...

Страница 46: ...ustralian plug 76199 01 5 7 Accessories Description Ohaus Part No Calibration Masses ASTM Class 1 Tolerance 20 g 49024 11 50 g 49054 11 100 g 49015 11 200 g 49025 11 500 g 49055 11 1 kg 49016 11 2 kg...

Страница 47: ...8 7x35 5 8 25 x 4 x 14 20 9 x 10 1 x 35 5 WxHxD in cm 5 8 Specifications Capacity g 62 110 162 210 100 210 Readability mg 0 1 0 1 1 Weighing units g mg oz oz t ct dwt taels 3 mommes gn ti N custom uni...

Страница 48: ...erature range 5 C to 40 C Atmospheric humidity 80 rh to 30 C Voltage fluctuations 15 10 Installation category II Pollution degree 2 Power supply voltage 12 V ac 50 60 Hz or 12 V dc 1A Power adapter AP...

Страница 49: ...43...

Страница 50: ...44...

Страница 51: ...ged by accident or misuse exposed to radioactive or corrosive materials has foreign material penetrating to the inside of the product or as a result of service or modification by other than Ohaus In l...

Страница 52: ...rights reserved 11780302B 0103 2 12 Ohaus Corporation OHAUS UK Ltd 19A Chapin Road 64 Boston Road P O Box 2033 Beaumont Leys Pine Brook NJ 07058 2033 USA Leicester LE4 1AW Tel 973 377 9000 England Fax...

Отзывы: