EXPLORER BALANCES
EN-103
NOTE:
The Information is dependent on the application and the ACCUMULATE setting.
Examples are shown below.
Mode
Accumulate
set
OFF
Accumulate
set
ON
Weighing
None
N: 10
Total: 10.000 kg
Avg: 1.000 kg
Std: 0.001 kg
Min: 0.999 kg
Max: 1.001 kg
Diff: 0.002 kg
Counting
APW: 0.100 kg
APW: 0.100 kg
N: 10
Total: 1000 Pcs
Avg: 100 Pcs
Std: 1 Pcs
Min: 99 Pcs
Max: 101 Pcs
Diff: 2 Pcs
Percent
Ref Wt: 1.23 kg
Ref Wt: 1.23 kg
Check
Weighing
Under: 0.995 kg
Over: 1.005 kg
Under: 0.995 kg
Over: 1.005 kg
N: 10
Total: 10.000 kg
Avg: 1.000 kg
Std: 0.001 kg
Min: 0.999 kg
Max: 1.001 kg
Diff: 0.002 kg
Dynamic
Level: 0
Level: 0
N: 10
Total: 10.000 kg
Avg: 1.000 kg
Std: 0.001 kg
Min: 0.999 kg
Max: 1.001 kg
Diff: 0.002 kg
5.9.10 Print Layout
This sub-menu is used to define the format of the data output to a printer or computer.
Format
Set the printing format.
MULTI
= a multiple line printout is generated. A CRLF is added after each data
output.
SINGLE
= a single line printout is generated. A TAB delimiter is added between
each data output.
Feed
Set the paper feed.
NONE
= the paper remains in its current position after printing.
LINE
= move the paper up one line after printing.
4 LINE
= move the paper up four lines after printing.
FORM
= a form feed is appended to the output.
5.9.11 Data Transfer Function
Output weighing results directly to a PC application.
Setup is easy and
no additional software is required.
OFF
= do not print.
ON
= print the specified settings.