
Defender Series Base EN-3
Troubleshooting
Aside from installing components and leveling adjustments, the Defender Series Base does not require any other
adjustments as shipped from the factory.
Operational difficulties that may be encountered can often be traced to simple causes such as:
Loose or incorrect wiring connections
Failure to remove the shipping spacers
Obstructions to the base frame
Unstable environments
Incorrect calibration or setup of the indicator
If the troubleshooting section does not resolve or describe your problem, contact your authorized OHAUS service
agent. For service assistance or technical support in the United States call toll-free 1-800-526-0659 between 8:00
AM and 5:00 PM EST. An OHAUS product service specialist will be available to provide assistance. Outside the
USA, please visit our web site, www.ohaus.com to locate the OHAUS office nearest you.
TECHNICAL DATA
Technical Specifications
The technical data is valid under the following ambient conditions:
Operating temperature: -10ºC to +40 ºC (14 °F to 104 °F)
Relative humidity: 100% relative humidity, non-condensing
Altitude: Up to 2,000m
Note:
Not all models may be available in your area. Check with your local OHAUS representative.
TABLE 2.1 SPECIFICATIONS
MODEL
i-D5K1S
i-D12K1R
i-D25K1R
i-D50K1L
i-D150K1L
Capacity
10 lb /
5 kg
25 lb /
12.5 kg
50 lb /
25 kg
100 lb /
50 kg
300 lb /
150 kg
Approved Resolution
NTEP/MC 5000
NTEP/MC 3000
Safe Overload Capacity
150% of capacity
Pan Dimensions
254 x 254 mm
10 x 10 in
305 x 305 mm
12 x 12 in
400 x 500 mm
15.7 x 19.6 in
Base Construction
316 stainless steel top and bottom frame,
hermetically sealed IP68/IP69K stainless
steel loadcell.
316 stainless steel removable
weighing platform with 316
stainless steel frame,
hermetically sealed
IP68/IP69K stainless steel
loadcell.
Repeatability
(std.deviation)
2d
Linearity
±2d
Load Cell Capacity
10 kg
20 kg
50 kg
100 kg
200 kg
Load Cell Cable
50cm L x 6-wire
1m L x 6-wire
Load Cell Type
350 Ohm, SST, single point
Load Cell Excitation
3.3-15V DC/AC
Load Cell Rated Output
2mV/V
Load Cell Protection
IP69K
Net Weight
6 kg / 13.2 lb
7.5 kg / 16.5 lb
16 kg / 35.3 lb
Shipping Weight
7 kg / 15.4 lb
8.5 kg / 18.7 lb
18 kg / 39.7 lb
Содержание Defender D5K1S
Страница 1: ...DefenderTM 6000 Series Base Instruction Manual...
Страница 2: ......
Страница 21: ......
Страница 22: ......
Страница 23: ......