background image

9

OHAUS CORPORATION

The rule is to lengthen the distance between the fulcrum and the load pivots to increase the indication. Conversely, shorten
the distance between the fulcrum and the load pivots to decrease the indication. Use hone part number 085061020 (fine) and
part number 085062020 (coarse) to adjust the pivots.

Note the direction of honing. Always hone away from the load edge of the pivot.

1. Side to Side

To correct a shift error side on a lever, note the indication at each location, and hone the load pivot as needed to correct
the side to side error on either lever. Hone only the load pivots to correct an error on either lever for a side to side correction.

2. Front to Back

When the indications on each load point of each lever are equal side to side and the indications of the levers are different
front to back, then hone the power pivot of the short lever until the two levers are equal front to back. Do not hone the
power pivot of the long lever. Increase the pivot distance from the power pivot to the fulcrum pivot on the short  lever
if the short lever has a higher indication than the log lever.

When the Shift error is corrected, recalibrate the scale and retest the shift. Continue calibration and shift test/adjustment
until no shift error is found after the scale is calibrated.

V.  ZERO ADJUSTMENT (as needed)

Check the zero balance by positioning the beam poise at "0" on the beam. Turn the locking lever to free the beam. The
beam should float without touching the top bottom of the trig opening. If it touched the top, turn the zero adjustment screw
(Figure 4) counter-clockwise with a screwdriver. Turn the screw clockwise until it is touching the bottom.

VI.  OPERATING INSTRUCTIONS

A.  SCALE  OPERATION

The OHAUS D500M beam utilizes a balancing beam principle to indicate weight. With the scale loaded, slide the
beam poise toward the trig loop (away from the column). Adjust the beam poise until the tip of the beam is in a balance
state within the trig loop. The position of the beam poise indication (red pointer) on the beam will indicate the amount
of weight on the scale platform.

NOTE: 

When operating the scale, use wedges at the two wheels to prevent the scale from moving.

TO LENGTHEN

THE DISTANCE

TO SHORTEN

THE DISTANCE

Содержание D500M

Страница 1: ...MODEL D500M PORTABLE BEAM SCALE ASSEMBLY MANUAL AND OPERATING INSTRUCTIONS OHAUS CORPORATION...

Страница 2: ...2 MODEL D500M PORTABLE BEAM SCALE ASSEMBLY MANUAL AND OPERATING INSTRUCTIONS...

Страница 3: ...I SPECIFICATIONS 04 III ASSEMBLY INSTALLATION 06 IV SHIFT 08 V ZEROADJUSTMENT 09 VI OPERATINGINSTRUCTIONS 09 OHAUS CORPORATION RESERVES THE RIGHT TO MAKE REFINEMENTS OR CHANGES WITHOUT NOTICE CAUTION...

Страница 4: ...lock secures the indicator when the scale is not in use This scale utilizes a rugged fabricated steel design and is equipped with phenolic wheels for easy portability A stainless steel platform is av...

Страница 5: ...5 OHAUS CORPORATION Figure 2 A Scale Platform B Spider Assembly C Beam Assembly D Column E Beam Shelf F Scale Base G Load Box Assembly H LevelBubble...

Страница 6: ...rapping from the scale base 2 Remove the scale platform from the scale base and set it aside 3 For reinstallation purppose note the location of the level bubble item H figure 2 on the spider assembly...

Страница 7: ...rod under the pivot at the end of the main lever of the scale base Raising the tip of the main lever and some slight twisting of the hook by hand may be required to get the hook under the pivot 12 Be...

Страница 8: ...uccessively If a correction is not needed proceed to the operating instructions In a mechanical base shift refers to the pivot distances of the levers If on a lever the distances are equal then a leve...

Страница 9: ...fulcrum pivot on the short lever if the short lever has a higher indication than the log lever When the Shift error is corrected recalibrate the scale and retest the shift Continue calibration and shi...

Страница 10: ...adding the sum of the additional counterpoise weights on the load box assembly to the amount indicated on the beam Example both additional 100 lb counterpoise weights are secured to the load box assem...

Страница 11: ...11 OHAUS CORPORATION...

Страница 12: ...2033 USA Phone 1 973 377 9000 Fax 1 973 944 7177 www ohaus com For information in other countries contact the manufaturer TOLEDO DO BRASIL IND STRIA DE BALAN AS LTDA R Manoel Cremonesi 1 Tel 55 11 435...