CHAPTER 4 TESTING
4.5.2 Adjusting Off Center Load
If the Off Center Load (Shift) is excessive, perform adjustment as follows:
Top view of Load Cell
Side view of Load Cell
3
2
1
4
Figure 4-2. Off Center Load adjustment points.
1. Place the test weight in the center of the Platform.
2. Tare the indicator.
3. Move the weight to point A and record the reading.
4. Move the weight to point B and record the reading.
5. Move the weight to point C and record the reading.
6. Move the weight to point D and record the reading.
7. If the reading at point A is negative, file at points 1 and 4 AT AN ANGLE.
8. If the reading at point B is negative, file at points 1 and 2 STRAIGHT ACROSS.
9. If the reading at point C is negative, file at points 2 and 3 AT AN ANGLE.
10. If the reading at point D is negative, file at points 3 and 4 STRAIGHT ACROSS.
Note:
It is not recommended that you try to adjust more than –5 counts if the beam has
been filed already. If the beam has not been filed previously, you can adjust –10 counts.
Remember, when filing you are weakening the beam. File a little at a time.
Ohaus Corporation www.ohaus.com
4-4
CKW Base Series Service Manual
Содержание CKW3R
Страница 2: ......
Страница 4: ......
Страница 6: ...TABLE OF CONTENTS CKW Base Series Scales Service Manual ii Ohaus Corporation www ohaus com...
Страница 12: ...CHAPTER 1 INTRODUCTION Ohaus Corporation www ohaus com 1 6 CKW Base Series Service Manual...
Страница 14: ...CHAPTER 2 DIAGNOSTIC GUIDE Ohaus Corporation www ohaus com 2 2 CKW Base Series Service Manual...
Страница 22: ...CHAPTER 4 TESTING Ohaus Corporation www ohaus com 4 6 CKW Base Series Service Manual...
Страница 26: ...CHAPTER 5 PARTS LISTS DIAGRAMS Ohaus Corporation www ohaus com 5 4 CKW Base Series Service Manual...
Страница 28: ...APPENDIX A GLOSSARY Ohaus Corporation www ohaus com A 2 CKW Base Series Service Manual...
Страница 29: ......
Страница 30: ...80252592 P N 80252592 SERVICE MANUAL CKW BASES...