CHAPTER 1 INTRODUCTION
1.8 HOW LOAD CELLS OPERATE
The difference in the output signal before and after the mass was placed on the platform is
taken, interpreted and displayed.
Figure 1-3. Route of electrical Load Cell input and output through Wheatstone Bridge circuit.
A Load Cell is a force sensor, which receives a voltage (excitation) from a regulated power
source in the Indicator, and sends back a low-voltage milli-volt (mV) relative to the force applied.
The Load Cell signal is read by the Indicator, which converts it to a numeric value and shows it
visually. This output value increases as weight is loaded to the Load Cell.
For example, a 500kg Load Cell with 2mV/V output and 5V excitation would have a change of
0.02mV per kg change in the load, that is, 10mV/500=0.02mV per kg. At a load of 250kg the
signal voltage would thus increase by 5mV from the value measured at no load.
Tension
Tension
Compression
Compression
+Excitation
(Supply)
–
Excitation
(Supply)
INPUT(volts)
LOADCELL
2 mV/V
Strain
Gauges
4 strain gauges
connected as a
Wheatstone
Bridge
OUTPUT (milli volts)
+
Sig
–
Sig
CKW Base Series Service Manual
1-5
Ohaus Corporation www.ohaus.com
Содержание CKW3R
Страница 2: ......
Страница 4: ......
Страница 6: ...TABLE OF CONTENTS CKW Base Series Scales Service Manual ii Ohaus Corporation www ohaus com...
Страница 12: ...CHAPTER 1 INTRODUCTION Ohaus Corporation www ohaus com 1 6 CKW Base Series Service Manual...
Страница 14: ...CHAPTER 2 DIAGNOSTIC GUIDE Ohaus Corporation www ohaus com 2 2 CKW Base Series Service Manual...
Страница 22: ...CHAPTER 4 TESTING Ohaus Corporation www ohaus com 4 6 CKW Base Series Service Manual...
Страница 26: ...CHAPTER 5 PARTS LISTS DIAGRAMS Ohaus Corporation www ohaus com 5 4 CKW Base Series Service Manual...
Страница 28: ...APPENDIX A GLOSSARY Ohaus Corporation www ohaus com A 2 CKW Base Series Service Manual...
Страница 29: ......
Страница 30: ...80252592 P N 80252592 SERVICE MANUAL CKW BASES...