background image

3-27

            CHAPTER 3  INDICATOR TESTING

3.12 CALIBRATION

When the CKW-55 Indicator is initially set up with a base, it must be calibrated before use to ensure
accurate weighing results.

Complete CKW Scale models are shipped pre-calibrated from the factory.  However, many conditions
may require that the scale be recalibrated.  Temperature changes, geographic gravity variations and
rough handling are few reasons why a scale may need recalibration.

NOTES

:

1. Make sure that appropriate calibration masses are available before beginning calibration.
2. Make sure that the scale base is level and stable during the entire calibration process.
3. Masses required to perform the procedures should be in compliance with the requirements of the

scale base being used with the Indicator.

4. Calibration is unavailable if LFT set to On.
5. Before calibration, allow the scale to warm up for approximately 5 minutes after stabilizing to room

temperature.

6. To cancel calibration, press the 

TARE 

button anytime during the process.  The Indicator will

retain previously stored calibration data.

3.12.1 Span Calibration

Span Calibration uses two points to adjust the scale.  One point is called the zero value where there is no
weight on the scale.  The other point is the Span value that is often, but not necessarily, at full capacity.
The CKW-55 Indicator allows the user to select alternate Span calibration values.  Span values closest
to the full scale capacity are recommended to give the best performance over the entire weighing range.
Other Span values may be specified due to limited availability of calibration masses or for improved
accuracy at a lower weight range.  Before calibrating the scale, check that the weighing base is correctly
level.

1. Press and hold the

 G/N/T 

Menu

  button until MENU is displayed.  When the

 G/N/T 

Menu

 

 button

is released, CAL is displayed.

2. With CAL displayed, press the 

Yes

 button, SPAN is displayed.

3.11 CALIBRATION USING AN EXTERNAL BASE (Cont.)

10.After wiring is completed and jumpers are in place, the Indicator can now be calibrated.

11.After testing, replace the cover.

Figure 3-13. Open and Shorted Jumpers.

Содержание CKW-55

Страница 1: ...Ohaus Corporation 19A Chapin Road P O Box 2033 Pine Brook NJ 07058 973 377 9000 CKW 55 INDICATOR SERVICE MANUAL T...

Страница 2: ......

Страница 3: ...sumes no liability arising from the use or misuse of this material Reproductionofthismaterialisstrictlyprohibited Material in this manual is subject to change Copyright 2006 Ohaus Corporation all righ...

Страница 4: ......

Страница 5: ...3 3 2 2 Simulator Testing 3 6 3 3 Power Supply Testing 3 7 3 3 1 Voltage measurements 3 7 3 4 Battery Testing 3 8 3 4 1 General Information 3 8 3 4 2 Troubleshooting the Battery 3 9 3 5 Recharge Board...

Страница 6: ...Housing 4 1 4 2 Replacing the Membrane Switch 4 1 4 3 Replacing the Main PC Board 4 2 4 4 Replacing the Power supply 4 2 4 5 Replacing the Optional Recharge Board 4 3 4 6 Replacing the Optional Relay...

Страница 7: ...4 2 7 Relay Operation Failure Optional board 2 4 2 8 Indicator Display Incorrect or Unstable 2 5 2 9 Unable to Maintain Time and Date Functions 2 5 2 10 Error Code Table 2 6 3 1 Membrane Switch Contin...

Страница 8: ...3 14 3 8 Relay Board Without Terminal Cover 3 14 3 9 RS422 485 Typical 4 wire Duplex Connection 3 18 3 10 RS422 485 Typical 2 Wire Half Duplex Connection 3 18 3 11 RS422 485 Location in Housing 3 19...

Страница 9: ...pter 4 RepairProcedures ContainsdetailedrepairproceduresforallmajorassembliesoftheCKW 55 Indicator including the optional Relay Board RS422 485 Communication Board the Recharge Board and Battery Chapt...

Страница 10: ...librationmassesrequiredisdependentonthecapacityandreadabilityofthebaseused Themasses required to test the CKW 55 Indicator when connected to an Ohaus CKW base must meet the requirements of ASTM Class...

Страница 11: ...om the AC power line pull out the plug The Indicator contains sensitive electronic components that are sensitive to electrostatic discharge We recommend that you wear a grounding wrist strap when doin...

Страница 12: ...1 4 CHAPTER 1 INTRODUCTION...

Страница 13: ...1 and observe and record power on information codes and software revision Record all menu settings if possible as received See section 3 9 4 On indicators equipped with an optional battery allow the...

Страница 14: ...repair performOperationalTestsinsection3 9 2 1 INDICATOR WILL NOT TURN ON WITH AC POWER Cont SYMPTOM Indicator will not turn on using battery power PROBABLE CAUSE Battery discharged or defec tive Wir...

Страница 15: ...Replaceifnecessary See Repair Procedure 4 2 If membrane switches are OK refer to section 3 2 and test the Main PC Board If defective replace the Main PC Board SeeRepairProcedures4 3 Afterrepair perfo...

Страница 16: ...on 3 9 TABLE 2 6 RS422 485 COMMUNICATIONS FAILURE OPTIONAL BOARD SYMPTOM Cannot communicate using RS232 PROBABLE CAUSE Menusettingsimproperlyset Main PC board defective REMEDY Refer to the Instruction...

Страница 17: ...nd date PROBABLE CAUSE Real Time Clock RTC bat tery depleted Main PC Board defective TABLE 2 9 UNABLE TO MAINTAIN TIME AND DATE FUNCTIONS REMEDY Replace the Real Time Clock battery Refer to section 4...

Страница 18: ...te the Indicator Clear platform of excess mass or recalibrate to establish a new zero point Go through all menus and re establish all settings or go to each menu and reset to default settings Check op...

Страница 19: ...al low battery to charge Refer to section 3 4 and test the battery Replace if necessary See Repair Proce dure 4 8 Clear platform of excess masses or recalibrate to establish a new zero point Refer to...

Страница 20: ...2 8 CHAPTER 2 DIAGNOSIS...

Страница 21: ...ld be replaced NOTE Depending upon pressure placed on the individual switch the resistance reading can vary anywhere from approximately 15 to 70 Ohms This is normal Check each switch on the panel The...

Страница 22: ...3 1 MEMBRANE SWITCH CONTINUITY TEST FUNCTION SWITCH MEASURE ACROSS PINS LIBRARY 2 11 NEXT 2 10 TARGET 3 10 ON ZERO 8 9 PRINT 4 11 G N T 4 10 TARE 3 11 1 5 10 2 5 11 3 5 12 4 6 10 5 6 11 6 6 12 7 7 10...

Страница 23: ...s Transistor Q1 and provides turn on of 12 V dc power to rest of circuit B IC U13 regulator provides 5 V dc to the board C IC U14 regulator provides the 5 V excitation for the external Load Cell D IC...

Страница 24: ...and is regulated If the voltage is low or is not present refer to section 3 3 and test the Power Supply Board If the Power Supply board is defective refer to section 4 4 and replace the board and con...

Страница 25: ...cell If this voltage is above or below these limits the Main PC Board should be replaced Refer to section 4 3 and replace the Main PC Board 9 Measure the voltage at IC U22 pin 2 location E it should...

Страница 26: ...sing refer to procedure 4 1 3 Disconnect the Load Cell cable from connector J2 on the Main PC Board Refer to Figure 3 3 for connector location 4 Connect the Simulator to connector J2 See Figure 3 2 fo...

Страница 27: ...se Check the fuse with an Ohmmeter If defective replace with a new fuse of the same type Reinstall protective fuse cover and reconnect ac power and measure voltage at connector CN2 If voltage is not p...

Страница 28: ...ssembly Battery Monitor Chip The battery assembly contains a monitor chip that monitors voltage and temperature The chip also contains a small amount of non volatile memory The number of charge cycles...

Страница 29: ...ttery operates the Indicator for less than 5 hours it should be replaced See section 4 8 for battery replacement procedures Note The power save function should be turned off to allow testing battery l...

Страница 30: ...he maintenance charge starts A red LED indicates the detection of a battery system failure If the battery is new and has never been charged a one time extended time trickle charge is started this one...

Страница 31: ...oard is working To confirm this momentarily press the C button at 1 or 2 minute intervals and note the battery voltage on the indicator display It should increase slightly with each button press As an...

Страница 32: ...n replaced allow the battery to charge several hours WARNING HAZARDOUS VOLTAGES ARE PRESENT WITH THE INDICATOR HOUSING OPENED AND AC POWER IS APPLIED USE CARE WHEN MAKING ALL MEASUREMENTS 3 5 3 Voltag...

Страница 33: ...s are replaceable The fuses one for each solid state relay output are replaceable 3 6 2 Relay Testing There are several reasons why the relays can fail they are listed as follows A Improper menu setti...

Страница 34: ...3 14 CHAPTER 3 INDICATOR TESTING Figure 3 7 Relay Board with Terminal Cover Figure 3 8 Relay Board without Terminal Cover 3 6 2 Relay Testing Cont...

Страница 35: ...er to the indicator Refer to the instruction manual and set the Indicator up to provide a test of each relay 3 For dc relay types use an Ohmmeter providing a minimum of 5 Volts DC and check each relay...

Страница 36: ...PC board voltage checks should be normal in accordance with section 3 2 NOTE A possible failure could have occurred on the Main PC board RS232 components and still have normal operating voltages pres...

Страница 37: ...ent Input Modes not used in range Mode xU Set weighing unit Where x 1 to 6 1 g 2 kg 3 lb 4 oz 5 lb ox 6 if x blank then increment to next available mode The unit can not be selected if it is not enabl...

Страница 38: ...e the jumper arrangement shown in Figure 3 10 and 3 11 RS422 485 Typical 4 Wire and Typical 2 Wire Connections Note W1 must be shorted to allow 2 wire communications W2 must be shorted on the last sca...

Страница 39: ...r and enter the COM menu and set the Baud Rate Parity Data Length Stop Bit Handshake Type to RS485 and address number then exit the menu 6 On the computer set the correct COM port and all settings on...

Страница 40: ...elay operation Stop Tare Start Stop T S S The first external input signal initiates a Tare function The second external input signal enables the output relay operation Start The third external input s...

Страница 41: ...and annunciators will light simul taneously at the previously selected setup brightness for 1 second Simultaneously the Beeper will sound at the previously selected setup volume for 1 2 sec Power On...

Страница 42: ...ion may be different depending on the last set up of the Indicator When Tare is not active Gross weight is displayed by default If Tare is active NET weight is displayed 3 9 2 Turning Off the Indicato...

Страница 43: ...WARNING HAZARDOUS VOLTAGES ARE PRESENT WITH THE INDICATOR HOUSING OPENED AND AC POWER IS APPLIED 1 Disconnect power from the Indicator Remove the line cord plug from the power source 2 Open the Front...

Страница 44: ...l OFF Display Signal OFF Graph Incr Auto Graph Display Bar Print Menu Print Stable ON Auto Print OFF Print Content On Date Time GMP User ID GMP Proj ID GMP Scale ID GMP Name ID Format Column Feed OFF...

Страница 45: ...ht 3 0 kg 6 61 lb Operating Temperature Range 10 C to 40 C 14 F to 104 F Power 100 240 VAC 50 60 Hz Universal Power Supply Hardwire Cord Optional Internal Rechargeable NiMH Battery 10 hour operation i...

Страница 46: ...ect the wires to Terminal Strip J2 according to the proper signal names See Figure 3 12 Tighten all terminal screws securely 5 When connecting a 2 mv v load cell Jumper W1 is to be left in place short...

Страница 47: ...ress the TARE button anytime during the process The Indicator will retain previously stored calibration data 3 12 1 SpanCalibration Span Calibration uses two points to adjust the scale One point is ca...

Страница 48: ...ation was successful the calibration weight is displayed and the scale exits automatically to the active weighing mode Remove the weight and begin weighing operations 3 12 2 LinearityCalibration Linea...

Страница 49: ...ndicator When any selection within the Read GMP menu is enabled calibration test results are automatically printed NOTE The Calibration Test function is available even when the CAL menu is locked out...

Страница 50: ...the Geo factor is set to correspond to the geographical location where the calibration is being performed Following calibration the Geo factor is changed to match the location where the scale is to b...

Страница 51: ...11 11 41 12 42 19 16 16 15 15 14 14 13 13 12 12 11 42 19 43 26 17 16 16 15 15 14 14 13 13 12 12 43 26 44 32 17 17 16 16 15 15 14 14 13 13 12 44 32 45 38 18 17 17 16 16 15 15 14 14 13 13 45 38 46 45 1...

Страница 52: ...s of troubleshooting and calibration 1 Remove power from the Indicator Refer to section 4 1 and open the Front Housing 2 See Figure 3 14 Printed Circuit Board Connections and remove the switch cover R...

Страница 53: ...and perform a span calibration 2 Follow the Indicator prompts When the Indicator indicates a given mass should be placed on the base set the Simulator to an equivalent value to simulate the output for...

Страница 54: ...DINGS FOR CKW BASES NOMINAL OUTPUT IN MILLIVOLTS AT 5 VOLTS EXCITATION CAPACITY ZERO 50 100 3 kg 4 3 2 0 Zero 2 3 Zero 4 7 6 kg 1 6 2 0 Zero 2 8 Zero 5 6 15 kg 1 7 2 0 Zero 2 4 Zero 4 8 30kg 1 3 2 0 Z...

Страница 55: ...fer to section 4 1 and open the Front Housing 3 With the Housing opened disconnect the connector from J1 on the Main PC board This is the cable from the Membrane Switch 4 Facing the front panel of the...

Страница 56: ...place and remove the Main PC board One screw contains a grounding clip 7 Replace the Main PC board and install the four screws previously removed Make sure the grounding clip is in place under one of...

Страница 57: ...r screws holding the Recharge board in place and remove the Recharge board 5 Replace the Recharge board and install the four screws previously removed 6 Reconnect connectors J1 and J2 to the Recharge...

Страница 58: ...ON BOARD The RS422 485 Communication board is supplied as an optional kit NOTE Read this procedure through first before proceeding 1 Remove AC power from the Indicator 2 Refer to section 4 1 and open...

Страница 59: ...ure 5 1 6 While holding the battery in place mount the center bracket using the supplied screws The mounting screws must remain loose at this step NOTE The center bracket is the part without the extra...

Страница 60: ...older 5 To install a new battery into the holder hold the battery with the mark facing up Slide the battery under the Contact Clip until the Retaining Clip springs up and locks the battery in place CA...

Страница 61: ...is made The Indicator MUST meets the parameters of all applicable specifications in this manual If further technical information is needed in the United States call toll free 1 800 526 0659 between 8...

Страница 62: ...5 2 CHAPTER 5 PARTS LISTS 5 1 CKW 55 INDICATOR EXPLODED VIEW Figure 5 1 Exploded View of CKW 55 Indicator...

Страница 63: ...g P7 CKW 7mm ID 17 71209067 I O Fitting P9 CKW 9mm ID 19 71209084 Cable Power Supply CKW 20 71209078 Fuse Power Supply Kit of 2 CKW 21 71209061 LFT Cover with hardware CKW 24 71154804 Gasket CKW 80251...

Страница 64: ...olumn TABLE 5 2 CKW 55 COLUMN PARTS LIST ITEM NO PART NO DESCRIPTION 1 71167972 Column Mount Accessory SS CKW Tube 230mm 1 80251059 Column Mount Accessory SS CKW Tube 6800 mm 2 71128010 Cable Clamp CK...

Страница 65: ...SV Enters the Monitor Mode esc RS Reset the scale similar to power off and on and exits the Monitor Mode esc EPX Indicator is set to 10 times expanded display mode X 1 ON X 0 OFF esc DF Resets all sca...

Страница 66: ...C 099 c 036 068 D 100 d 037 069 E 101 e 038 070 F 102 f 039 071 G 103 g 040 072 H 104 h 041 073 I 105 i 042 074 J 106 j 043 075 K 107 k 044 076 L 108 l 045 077 M 109 m 046 078 N 110 n 047 079 O 111 o...

Страница 67: ......

Страница 68: ...19A Chapin Road P O Box 2033 Pine Brook NJ 07058 USA Tel 973 377 9000 Fax 973 944 7177 www ohaus com With offices worldwide www ohaus com P N 80251063 Ohaus Corporation 2006 all rights reserved Printe...

Отзывы: