Load Cell Replacement

Remove the scale platform and disconnect the

indicator battery and AC power source.

IMPORTANT NOTE

Wait 30 seconds after removing power to the indicator before unplugging the load

cell cable.

Remove the top load cell mounting bolts that secure the top frame to the load cell.

Set the top frame and load cell spacer aside.Remove the bottom load cell

mounting bolts. The load cell assembly can be removed from the lower base.

When reinstalling a load cell, reverse the preceding steps. Lubricate the threads

and under the head of the hex bolts before reinstalling. Using a torque wrench,

tighten the hex bolts to the following torque specifications:

Model

Torque

CH15R

20 N.m (15 Ft/Ib)

CH30R

20 N.m (15 Ft/Ib)

CH60R

20 N.m (15 Ft/Ib)

CH60L

20 N.m (15 Ft/Ib)

CH100R

30 N.m (22 Ft/Ib)

CH150R

40 N.m (30 Ft/Ib)

CH300R

50 N.m (37 Ft/Ib)

Overload Stop Adjustments

The overload stop gaps must be checked and reset if the top or

bottom frame, or load cell is replaced. To set the gaps, remove

the platform, loosen the jam nuts (refer to Figure 1-1), then use

the proper size feeder gauge in the gap, turn the set screws

until a slight drag on the feeder is felt. Tighten the jam nut and

recheck the gap. Re-adjust if necessary, cover the platform

and check for full capacity. Refer to Figure 1-2 for the location

of the overload stops and Table A for the gap settings per

Order Number.

Содержание CHAMP II BASE

Страница 1: ...Ohaus Corporation 29 Hanover Road P O Box 900 Florham Park NJ 07932 0900 Phone 973 377 9000 Fax 973 593 0359 CHAMP II SCALE with CD11 Operating and Installation Instructions...

Страница 2: ...9 336 92 31 93 68 Compatibilit elettromagnetica EC1010 1 EN60950 1992 Safety Regulations IEC1010 1 EN60950 1992 Sicherheitsbestimmungen IEC1010 1 EN60950 1992 Consignes de s curit IEC1010 1 EN60950 19...

Страница 3: ...MFUL INTERFERENCE IN WHICH CASE THE USER WILL BE REQUIRED TO CORRECT THE INTERFERENCE AT HIS OWN EXPENSE THIS DIGITAL APPARATUS DOES NOT EXCEED THE CLASS A LIMITS FOR RADIO NOISE EMISSIONS FROM DIGITA...

Страница 4: ...Capacity Ultimate Overload 300 of Full Capacity Operation Environment Champ II is designed to meet NTEP 3000d requirements to operate in a temperature range from 10 C to 40 C 0 to 95 relative humidity...

Страница 5: ...REMOVE THE RED PLASTIC SHIPPING TABS IN THE CORNERS OF THE SCALE BASE Troubleshooting If operational difficulties are encountered first obtain as much information as possible regarding the problem Fai...

Страница 6: ...ng a torque wrench tighten the hex bolts to the following torque specifications Model Torque CH15R 20 N m 15 Ft Ib CH30R 20 N m 15 Ft Ib CH60R 20 N m 15 Ft Ib CH60L 20 N m 15 Ft Ib CH100R 30 N m 22 Ft...

Страница 7: ...platform weigh within tolerance If the scale does not pass the shift test verify the overload stop gaps before replacing the load cell No adjustment for the shift is possible If the shift test cannot...

Страница 8: ...he field before use Parts list OHAUS PART NO DESCRIPTION 71129642 Indicator Bracket Assembly 71125105 Stainless Steel Column 350 mm high 71125103 Stainless Steel Column 680 mm high 71129064 Column Bra...

Страница 9: ...9...

Страница 10: ...e switches Power 9 VDC 150 mA AC adapter or six alkaline C type batteries Typical battery life 250 hours 100 hours continues use Weighing units g kg lb Modes Weighing parts counting Zeroing Range 2 18...

Страница 11: ...enetrating to the inside of the product or as a result of service or modification by other than Ohaus In lieu of a properly returned warranty registration card the warranty period shall begin on the d...

Отзывы: