2
CD-1 Indicator
2.2 Selecting the Location
The CD-1 Indicator should be used in an environment which is free from vibration, temperature or humidity extremes.
These factors will affect displayed weight readings. Scale bases used with the CD-1 Indicator should be located on a
stable, level surface and kept away from vibrating sources such as large machinery. Maximum accuracy will be
achieved when the working area is clean and vibration free.
2.3 Cautionary Notes
The CD-1 Indicator
must not be operated in wet or hazardous areas.
Before connecting the AC adapter, verify that the indicated voltage corresponds to the local mains voltage.
2.4 Connecting the CD-1 Indicator to a Scale Base
The scale base is connected to the round, 4-pin connector located on the right side of the CD-1 Indicator. The mating
connector supplied in the packaging must be attached to the load cell cable. Slide the connector shell over the load cell
cable. Orient as shown in the illustration. Refer to the scale base manufacturer's information to determine the proper
load cell connections. Solder the load cell cable to the connector contacts according to table 2-1 and refer to the the
connector illustration.
NOTE
: Any extra load cell connection wires should be isolated and insulated.
TABLE 2-1. LOAD CELL CONNECTIONS.
Connector Pin No.
Connection
1
- Excitation
2
- Signal
3
+ Signal
4
+ Excitation
2.5 Connecting Power
The CD-1 Indicator may be operated using the supplied AC Adapter, or with 6 Alkaline "AA" batteries (not supplied).
2.5.1 AC Adapter
Connect the AC Adapter connector to the receptacle located at the right-hand side of the CD-1 Indicator and
plug the adapter into a convenient outlet.
NOTICE:
The socket/outlet must be installed near
the equipment and shall be easily accessible.
2.5.2 Battery Installation
Open the battery cover on the bottom of the housing.
Insert 6 Alkaline "AA" batteries into position. Orient the batteries so that the positive (+) ends rest against the
reeds and the negative (-) ends against the springs.
NOTE
: When the CD-1 Indicator is operated from batteries, the Auto Shut-Off feature is automatically enabled to
extend battery life. When no activity occurs within a 5 minute period, the CD-1 Indicator shuts off. This
feature cannot be disabled.
AC ADAPTER CORD
4
3
1
2
CONNECTOR REAR VIEW
After soldering the wires to the contacts, thread the shell onto the receptacle. Tighten the strain relief clamp over the
load cell cable. Plug the connector into the mating connector on the Indicator and tighten the knurled nut.
STRAIN RELIEF
CLAMP
KNURLED
NUT
SHELL
CABLE
RECEPTACLE
CONTACTS
CONNECTOR ASSEMBLY
Содержание CD-1
Страница 2: ......
Страница 15: ......
Страница 17: ...MANUAL DE INSTRUCCIONES Indicador modelo CD 1...
Страница 18: ......
Страница 31: ......
Страница 33: ...MANUEL DE L UTILISATEUR Indicateur mod le CD 1...
Страница 34: ......
Страница 47: ......