ADV Service Manual P/N 83020790 A
Ohaus Corporation
www.ohaus.com
22
2. Locate the four screws holding the Transducer Shield in place. Remove the screws.
3. Remove the Transducer Shield. The Transducer is now exposed for adjustments.
4. Reinstall the Platform on 1mg balance and Sub Platform and Weighing Platform on 10/100mg balance onto
the exposed Transducer shaft.
NOTE
:
Do not attempt these adjustments on the balance unless the balance is free
from drafts and is level. The balance is very sensitive and adjustments will
be affected.
5. Plug the power cord into the balance and turn the balance on.
6. Place 1/2 of the balance capacity or greater on the rear of the Platform.
7. Tare the balance. Move mass to right front and left front of the Platform as shown in Figure 4-3 and note
the readings. Correct the worst error first by turning Adjusting Screws on the Transducer shown in Figure 4-4.
Figure 4-3. Mass Placement Locations.
Figure 4-4 Load Adjustments Screw Locations
8. Repeat the Off-Center Load Test and adjustments until balance is within specifications.