EN-14
Adventurer Pro
3.5.3 Percent Weighing (Cont.)
Place the sample on the pan and press the
Function
button. The example illustrates 50
grams as the reference weight.
Remove the reference weight and place the item
to be compared on the pan. The balance
indicates the actual percentage on the primary
display. The secondary display indicates the
weight in the selected measuring unit.
Pressing the
Function
button momentarily
displays the reference weight on the secondary
display.
Clearing the Reference Weight
Press and hold the
Mode
button until PERCENT
is shown on the secondary display.
3.5.4 Check Weighing
Enter this mode to compare the weight of a sample against target limits. Refer to paragraph
3.4.3 to enable Check Weighing.
USE
The secondary display shows EDIT REF?
(blinking) when limits exist. Press
No
to use
existing limits and begin Check Weighing. If
no limits are stored, or if EDIT REF is chosen,
Under and Over limits must be set.
Setting Under and Over Limits
The primary display shows UNDER or OVER
indicating which limit is being entered. The
secondary display shows the previous setting
or all zeros if no limit was stored (blinking).
Press the
Yes
button to accept this setting.
Press the
No
button to edit this setting. The
Under limit is the first to be edited, followed by
the Over limit.
3 Adventurer Pro Text pages 3 to 14.P65
4/10/2003, 6:00 AM
14
All manuals and user guides at all-guides.com