12
No gas charging
in the seat-pad.
①
The terminal of the snuffle valves and
wires are poorly connected.
②
The snuffle valves fail.
③
The inflator pump fails.
④
Main PCB fails.
①
Plug the terminal securely.
②
Replace the snuffle valves.
③
Replace the inflator pump.
④
Replace main PCB.
13
No gas charging
in the foot rest.
①
The terminal of the snuffle valves and
wires are poorly connected.
②
The snuffle valves fail.
③
The inflator pump fails.
④
Main PCB fails.
①
Plug the terminal securely.
②
Replace the snuffle valves.
③
Replace the inflator pump.
④
Replace main PCB.
Содержание OG8020