OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ofite.com
29
Onboard
Display
Test Setup - Profile
On the Profile tab of the Test Setup screen, you can create a Test Profile for
your test.
1. Choose a Parameter (Temperature, Pressure, or Motor).
2. Choose a Step Type:
Temperature and Pressure
Ramp:
This will increase the temperature or pressure up to the target in a
set number of minutes. Enter the ramp time and target.
Step:
This will increase the temperature or pressure up to the target as
fast as possible. Enter the target temperature or pressure.
Dwell:
This will hold the current temperature or pressure for a set number
of minutes. Enter the time.
Motor:
On or Off
3. Enter the parameters for the step (ramp time, target temperature, motor
speed, etc).
4. Select the “ADD STEP” button.
5. To remove a step, select it in the list and select the “REMOVE STEP” but-
ton.
6. When you are finished adding steps, select the “SAVE” button.