background image

OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ofite.com

4

Parts and 

Accessories

Caps

#175-13-02  Outer Cap, 303 Stainless Steel

For tests below 400°F:

#175-18-02  Inner Cap with Pressure Port, 303 Stainless Steel

#175-18-03  Inner Cap with Pressure Port, 316 Stainless Steel

#175-65-002  Inner Cap without Pressure Port, 316 Stainless Steel

For tests above 400°F:

#175-00-3 

Inner Cap, Accepts Rupture Disk, 316 Stainless Steel

#175-56 

Rupture Disk, ¼", 2,000 psi

#175-57 

Rupture Disk, ¼", 1,500 psi

For tests above 400°F, always use an inner cap with a rupture disk.

O-rings and Gaskets

For tests below 200°F:

#175-54 

Buna N O-ring for Outside of Aging Cell

For tests up to 400°F:

#170-17 Viton

®

 O-ring for Valve Stem

#175-09-2 Viton

®

 O-ring for Inside of Aging Cell Body

#175-47 Viton

®

 O-ring for Outside of Aging Cell

#175-62 Viton

®

 O-ring for Teflon

®

 Liner Plug

#175-63 Viton

®

 O-ring for Teflon

®

 Liner Lid (Piston)

For tests above 400°F:

#175-09-1  Teflon

®

 O-ring for Inside of Aging Cell Body

#175-46 

Teflon

®

 O-ring for Outside of Aging Cell

#175-03 

Peek O-ring for Inside of Aging Cell Body

Buna N O-rings should only be used for temperatures below 200°F.

Viton

®

 O-rings can be used at temperatures up to 400°F.

Teflon

®

 O-rings should be used for temperatures above 400°F.

Set Screws and Wrenches

#175-14  Set Screw for Pressurized Aging Cells, ⅜"

#175-15  Wrench for ⅜" Set Screw

#175-60   

Teflon

®

 Liner, for 500 mL Aging Cells with T-screw

#175-60-1 Liner

#175-60-2 Piston

#175-60-3 Plug

#175-60-4 T-screw

#175-62 Viton

®

 O-ring for Teflon

®

 Liner Plug

#175-63 Viton

®

 O-ring for Teflon

®

 Liner Piston

Содержание 175-65

Страница 1: ...0 mL Double Capped 175 65 Instruction Manual Updated 6 18 2019 Ver 1 OFI Testing Equipment Inc 11302 Steeplecrest Dr Houston Texas 77065 U S A Tele 832 320 7300 Fax 713 880 9886 www ofite com Copyrigh...

Страница 2: ...lecrest Dr Houston TX 77065 USA Tel 832 320 7300 Fax 713 880 9886 www ofite com 1 Intro 2 Components 3 Parts and Accessories 4 Specifications 5 Operation 6 Maintenance 8 Warranty and Return Policy 9 T...

Страница 3: ...perature testing up to 500 F 260 C For prolonged exposure to elevated salinity at high temperatures such as 20 000 mg L chlorides at 350 F 176 6 C cells con structed from premium metals such as Hastel...

Страница 4: ...Wrench for Set Screw 175 16 Valve Stem 175 47 O ring for Outside of Aging Cell Body Viton 90 Qty 2 175 65 001 Cell Body 316 Stainless Steel 175 65 002 Inner Cap without Pressure Port 316 Stainless Ste...

Страница 5: ...o 400 F 170 17 Viton O ring for Valve Stem 175 09 2 Viton O ring for Inside of Aging Cell Body 175 47 Viton O ring for Outside of Aging Cell 175 62 Viton O ring for Teflon Liner Plug 175 63 Viton O ri...

Страница 6: ...e outer perimeter on the top and bottom of the cells Failure to do so can damage the rollers in the oven Teflon 175 46 Viton 175 47 and Buna N 175 54 o rings are available Cell Corrosion Test fluids u...

Страница 7: ...o ring so that it seats in place Hand tighten the outer cap in place Using the Allen Wrench tighten the set screws in the outer cap The ends of the cell body are identical The inner cap without the pr...

Страница 8: ...m the valve stem or cell cap 7 Place the Aging Cell inside the oven and adjust to the desired tempera ture 8 After the desired aging time has elapsed remove the cell from the oven and allow it to air...

Страница 9: ...se to the threads on the outer cap and cell body Every three years yearly for highly corrosive conditions 1 Sand blast the metal components cell body inner cap outer cap 2 Visually inspect the cell bo...

Страница 10: ...ny losses or damages that occur as a result of Improper installation or maintenance of the products Misuse Neglect Adjustment by non authorized sources Improper environment Excessive or inadequate hea...

Отзывы: