OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ofite.com
33
To replace the Seat Assembly:
1.
Unscrew the spring case (housing cover). This may require a strap wrench.
2.
Remove the spring case.
3. Remove the rubber diaphragm (#143-00-1) from inside the spring case.
4. Remove the thrust plate (#143-00-7).
5.
Using a wrench, loosen and remove the brass colored retainer.
6. The thrust pin and seat are now exposed and can be removed from the base.
7.
Replace the spring – repair kit.
8. Replace the seat holder – repair kit.
9. Replace the thrust pin and the Teflon seat which can be attached before
insertion-repair kit.
10.
Replace the brass retainer and tighten with a wrench.
11.
Replace the thrust plate with the curved edges downward.
12.
Place a new rubber diaphragm inside the Soring case
13. Replace the spring case and hand tighten.
Power Cord Maintenance
1.
Check power cord for Insulation wear and loose connections near the
heating jacket and plug.
2.
The heating jacket should be placed no farther than the distance of the power
cord from an electrical outlet.
3. The Power Cord should be kept away from the hot surface of the heating
jacket, while in use.
Any customer installed wiring, power cords or electrical connectors will Void
All Warranties.
Storage Space
1.
Consumables and tools should all be kept in one dedicated place.
2.
Components such as manifold assemblies should be kept in a similar
dedicated place.