25. Open (loosen) the top valve stem 

½

 turn to pressurize the sample. Main

-

tain this pressure on the fluid until the temperature has stabilized.

26. After the one hour heat up time, increase the pressure on the top pres

-

sure unit to 500 PSI (3,448 kPa) more than the back pressure. 

27. Open (loosen) the bottom valve stem 

½

 turn to initiate filtration as soon as 

possible. 

Closely monitor the pressure gauges. If at any time during the test the 

pressure inside the cell rises above the setpoint, carefully open the 

needle valve on the top pressure assembly just enough to bleed off the 

excess pressure. Then close the valve. If the cell pressure decreases due 

to collection of filtrate, increase the pressure with the inlet regulator. 

28. To collect filtrate, carefully open the ball valve on the bottom of the back 

pressure receiver while holding a graduated cylinder up to the valve port. 

Close the valve immediately after the pressure begins to decrease and 

the filtrate is collected. 

Collect filtrate at 10 seconds, 1 minute, 7.5 minutes, and 30 minutes. 

The initial 10 second collection is precautionary in nature, as a fluid with 

little filtration properties may fill up the 15 mL receiver tube almost imme

-

diately, potentially damaging the regulator. Do not record the 10 second 

collection as a separate notation, but do record the volume collected at 

the other time intervals. 

While collecting filtrate, maintain the test temperature within ± 5°F (±3°C). 

If the back pressure rises during the test, cautiously reduce the pressure 

by opening the valve on the receiver and drawing off some of the filtrate 

into the graduated cylinder.

29. 

At the end of the test, close (tighten) the top and bottom valve stems to 

seal off the cell.

30. Turn both regulator T-screws counterclockwise to close off the flow of 

pressurized gas.

31. Open the outlet valve on the back pressure receiver to collect all of the 

remaining filtrate.

32. Open the bleeder valve on the manifold block to release any remaining 

pressure in the line.

33. Remove the manifold block and the back pressure receiver. Drain any 

residual filtrate from the receiver into the graduated cylinder.

The heating jacket will still be very hot. Be careful not to touch it while 

removing the back pressure receiver. 

Содержание 170-182-1

Страница 1: ...ped Test Cell and N2 Pressuring Manifold 170 182 115 V 170 182 1 230 V Instruction Manual Updated 6 20 2017 Ver 3 0 OFI Testing Equipment Inc 11302 Steeplecrest Dr Houston Texas 77065 U S A Tele 832 3...

Страница 2: ...USA Tel 832 320 7300 Fax 713 880 9886 www ofite com 1 Intro 2 Specifications 3 Components 4 Safety 8 Quick Start 12 Operation 15 Data 20 Cell Cap Assembly 21 Maintenance 22 Diagrams 24 Test Cell 25 Wa...

Страница 3: ...OFITE Filtration devices fully conform to American Petroleum Institute API specifications A complete HTHP Filter Press consists of a controlled pressure source usu ally Nitrogen pressurization or Car...

Страница 4: ...Shipping Weight 38 lbs 17 2 kg Maximum Temperature Heating Jacket 400 F 204 C Maximum Temperature Cell 500 F 260 C Maximum Pressure Cell 5 000 PSI Maximum Pressure Receiver 750 PSI 5 1 MPa Pressure So...

Страница 5: ...15 mL Stainless Steel Tube for N2 144 11 90 Street Ell 144 15 Plate Brass Bushing NPT Male to NPT Female 170 07 O ring 170 28 Receiver Body 170 32 NPT Male Needle Valve 171 23 1 Safety Pin with Lanya...

Страница 6: ...71 38 1 000 PSI Gauge 2 NPT Bottom 171 40 1 500 PSI Gauge 2 NPT Bottom 171 42 3 000 PSI Gauge 2 NPT Bottom 171 90 06 MNPT FNPT Reducing Bushing 316 Stainless Optional 170 13 O ring for Test Cell Buna...

Страница 7: ...nician to perform a more uniform and reproduc ible test while maintaining a high degree of safety As optional items the us age is not compulsory but consideration should be given to these items when r...

Страница 8: ...eplecrest Dr Houston TX 77065 USA Tel 832 320 7300 Fax 713 880 9886 www ofite com 7 Thermocouple Assembly 171 45 1 Direct temperature measurement Of the fluid Inside the Cell Stand for HTHP Cell Assem...

Страница 9: ...eceiver is not required as the filtrate will not reach the boiling point However when operat ing above the boiling point of water a suitable Back Pressure Receiver is required otherwise the test fluid...

Страница 10: ...eful not to touch the outside of the jacket at any time during the heat up period or the test It is especially easy to get burned when installing or removing the Back Pressure Receiver as the technici...

Страница 11: ...ng in a small confined space Safety Retainer Pin Always use a Safety Retainer Pin with an attached lanyard and get in the habit of securing the cotter pin to the pin Improper placement of retainer pin...

Страница 12: ...d should be used for tests up to 400 F only The second set is made from Viton 90D OFITE suggests that all o rings cell caps cell bodies and valve stem are changed each time after running a test of 350...

Страница 13: ...cell cap into the cell and tighten it completely 11 Screw a valve stem into the inlet cell cap and tighten it completely 12 Pour the test fluid into the cell Leave adequate space for fluid expansion...

Страница 14: ...e both valve stems 29 Turn both regulator T screws counterclockwise 30 Open the outlet valve on the back pressure receiver to collect any remain ing filtrate 31 Open the valve on the manifold block to...

Страница 15: ...4 Manifold Block 170 20 Heating Jacket 171 00 or 171 01 Thermostat 171 71 Thermometer 154 10 Back Pressure Receiver 171 10 Nitrogen Manifold 171 24 Valve Stem 170 16 Regulators Nitrogen Connection Mal...

Страница 16: ...y Discard any components that show signs of corrosion damage or wear 4 Inspect all o rings Replace any that show signs of damage or wear Looks for nicks cuts or brittle o rings Replace all o rings aft...

Страница 17: ...id 13 Invert the cell body and carefully pour the sample into the cell Leave 2 5 1 cm void space for thermal expansion Be careful not to spill fluid on the o ring shoulder inside the cell 14 Place an...

Страница 18: ...nlet on the pressure manifold Open the pressure release valve on the Nitrogen bottle and note the bottle pressure as registered on the middle manifold gauge 24 Keeping the valve stems closed adjust th...

Страница 19: ...ted Collect filtrate at 10 seconds 1 minute 7 5 minutes and 30 minutes The initial 10 second collection is precautionary in nature as a fluid with little filtration properties may fill up the 15 mL re...

Страница 20: ...l body Pressure cannot be relieved from the cell by opening the outlet valve stem as the filter cake will seal off the cell It is a good idea to open the valve stem with the cell inside a sink or with...

Страница 21: ...is the amount of filtrate collected before the filter cake has had a chance to form and is expressed in millimeters To calculate the spurt loss use the following equation V1 2 V7 5 V30 V7 5 2 2V7 5 V...

Страница 22: ...d the cap 3 Place the retaining ring 130 81 040 into the groove around the outside of the cap Make sure it engages completely around the circle The cap should turn freely inside the locking ring 4 Pla...

Страница 23: ...old separately To replace the seat assembly a Unscrew the spring case housing cover This may require a strap wrench b Remove the spring case c Remove the rubber diaphragm from inside the spring case d...

Страница 24: ...or inlet filter Repair the regulator Gas leak from any pipe thread joint Loose fitting Remove the connec tion Clean the affected surfaces Reapply Teflon tape and tighten Gas leak from relief valve Fau...

Страница 25: ...e 170 32 Manifold Block 170 20 Safety Pin 171 23 1 With Lanyard Outlet Valve Stem 170 16 Nitrogen Manifold Primary Pressure Gauge 171 40 Source Pressure Gauge 171 42 Back Pressure Gauge 171 38 Connect...

Страница 26: ...ap 171 190 033 S Cell Body 171 190 020 S Retaining Ring 130 81 040 Outlet End Cap 171 190 032 S Cap Screw 171 190 023 Detachable Screen Mesh 170 18 O ring 228 Viton 90D 171 190 060 NPT Rupture Plug 5...

Страница 27: ...y losses or damages that occur as a result of Improper installation or maintenance of the products Misuse Neglect Adjustment by non authorized sources Improper environment Excessive or inadequate heat...

Отзывы: