Corrective action
Check the motor connection cable and connect cor-
rectly; replace if necessary.
Replace the pump.
Balance the rotor (only by the OLV Service
Department).
Bearings will have to be replaced (only by the OLV
Service Department).
Change the masses of the system or install vibration
damper to isolate oscillations.
Check the measurement device.
Clean or replace the measurement gauges.
Locate the leaks.
Bake out the pump; see Section 5.3.
Have the pump cleaned (only by the OLV Service
Department).
Check ultimate pressure of the forevacuum pump or
install a more powerful forevacuum pump.
Locate and repair leaks (only by the OLV Service
Department).
Check the connector lines; interchange poles if
necessary.
Check the forevacuum pump; install a more powerful
forevacuum pump if necessary.
Seal leak; install a more powerful forevacuum pump if
necessary.
Ensure sufficient supply of cooling air.
Route cooler air to the fan or employ water
cooling option.
Ensure sufficient supply of cooling water.
Have the pump repaired (only by the OLV Service
Department).
Repair or replace the forevacuum pump.
Check the airing valve and replace if indicated. Air the
TURBOVAC correctly; see Section 4.6.
Install a roughing line or pre-pump for a shorter period
of time or install a sorption trap.
Regenerate or replace the sorption trap.
Possible cause
Motor connection cable not attached, is
loose or is defective.
Pump has seized.
Rotor is out of balance.
Bearing is defective.
Pump running within the natural frequency
range of the system, causing resonance.
Measurement device is defective.
Measurement gauges are soiled.
Leak at the system, lines or pump.
Minor grime collection at the pump.
The pump is oily.
Forevacuum pump with insufficient pum-
ping speed or ultimate pressure which is
too high.
Leak at the power cord passage port.
TURBOVAC is rotating in the wrong direc-
tion.
Forevacuum pressure too high.
Gas volume too great / leak in the system.
Ventilation unit blocked.
Ambient temperature is too high.
Cooling water is lacking or insufficient.
Bearings are defective.
Anti-suckback valve at the forevacuum
pump is defective.
The TURBOVAC was not aired or impro-
perly aired when shut down.
System configured incorrectly: oil vapor
streams back during forepump operation.
Sorption trap is saturated.
Malfunction
TURBOVAC does not
start.
TURBOVAC generates
loud running noises and
vibrations.
The TURBOVAC does
not achieve ultimate
pressure.
TURBOVAC overheats
(malfunction indication at
the TURBOTRONIK).
The TURBOVAC or the
vacuum chamber is con-
taminated with oil.
Troubleshooting
32