Installation
GA10218_002_A7 - 9/2015 - © Oerlikon Leybold
29
Only 3-core mains cables having a protection ground conductor must be used.
Operation of the PHOENIX L300i where the ground conductor has been left
unconnected is not permissible. The PHOENIX L300i can be damaged when using
the wrong voltage. The voltage must be in the range 230V (+/- 5%), 115V (+/- 5%)
or 100V (+/- 5%) depending on the version.
The mains voltage rating for the PHOENIX L300i can be read off from the name
plate beneath the mains socket Fig. 9/4 at the side. This voltage is fixed and can
not be changed.
A separate fuse for each of the mains conductors has been integrated into the
mains switch.
The mains voltage is applied to the instrument via the detachable mains cable
which is supplied with the instrument. A mains socket Fig. 9/4 is available for this
purpose at the side of the instrument.