6.5 Parameter
Mode
6.5.1 Selection
The selection of the parameters is made by the configuration switch (
)
Figure 3 D,
page 12) on rear of the DISPLAY ONE. Proceed in addition as follows:
•
Switch the instrument off
•
Set the DIP switches as requiered
•
Switch the unit on once more
IMPORTANT INFORMATION:
When changing the DIP switch settings while the instrument is on, this will
initially have no effect. Any changes to the DIP switch settings will only come
into effect after switching the instrument on once more. DIP switch 4 is
intended for service functions and must always remain in the position OFF.
6.5.2 Description
Over the configuration switch with DIP switches 1 and 2 the mass unit is stopped as well
as selected with DIP switch 3 between two or three-figure measured value announcement.
6.5.3 Mass
unit
The mass unit is stopped over the DIP switches 1 and 2 of the configuration switch.
Mass unit
DIP switch 1
DIP switch 2
mbar
OFF
OFF
Pa
ON
OFF
Torr
OFF
ON
Table 6
–
Parameter Mass unit
•
Configure the mass unit according to the Table 6 on page 16 .
6.5.4 Display
Format
The display format is stopped over the DIP switch 3 of the configuration switch.
Display Format
DIP switch 3
3-stellig
OFF
2-stellig
ON
Table 7
–
Parameter Display Format
•
Configure the display format according to the