OEM PCA-20 Скачать руководство пользователя страница 8

 

 

Installation 

 

 

Only qualified personnel is allowed to install and perform the connection of PCA-20 
(PCA-25) into customer’s system 

 

PCA-20 (PCA-25) is intended only for capacitor charging applications. The module 
should only be used inside the medical equipment, which has means to isolate its circuits 
electrically from the supply mains on all poles simultaneously 

 

Use four M4 mounting holes on the bottom side of PCA-20 (PCA-25) for fixation of the 
module in your system. Refer to the 

Dimensional drawing

 section for their location. 

Note: max. tightening depth of mounting screws should not exceed 4mm

 

 

Connect the module to your system protective ground using provided M4 grounding 
thread. 

Note: max. screw depth should not exceed 6mm

 

 

Connect INPUT, HV OUTPUT and INTERFACE cables according to their functions 

 

Since module is air cooled, at least 50mm of free space should be provided both at the 
suction and pressure sides of PCA-20 (PCA-25) 

 

Operations 

 

1.

 

Disable

 

the high voltage output using 

INHIBIT

 signal (

PIN7 

of 

INTERFACE

2.

 

Set the desired output voltage in working range (0V-V

MAX

) by applying a DC voltage  

to the 

Voltage Program 

(

PIN3 

of 

INTERFACE

3.

 

Apply power to the module, wait 2-5s until module starts 

4.

 

Enable

 

the high voltage output (

PIN7 

of 

INTERFACE

5.

 

After operations disable the high voltage output,  
then disconnect module from the mains 

 

Faults 

 

Fault 

Cause of the fault 

Troubleshooting 

Overvoltage (Pin 19, pin 14) 

Voltage on the load exceeds 
110% of V

MAX

 level 

The fault signal will disappear after 
elimination of its cause and resetting 
INHIBIT signal (inhibit and enable the 
signal referring to the INTERFACE 
description) 

Short-circuit (Pin 19) 

HV OUTPUT is enabled, but 
voltage doesn’t increase in 
predefined time (1.5-2.0s) 

The fault signal will disappear after 
elimination of its cause and resetting 
INHIBIT signal (inhibit and enable the 
signal referring to the INTERFACE 
description) 

Open-circuit (Pin 19) 

There is no capacitive load 
connected to HV OUTPUT 

The fault signal will disappear after 
elimination of its cause and resetting 
(turning off and on) mains supply of the 
charger 

Overtemp (Pin 13) 

Temperature of internal elements 
exceeds the preset limit (approx. 
80°C) 

The fault signal will disappear automatically 
after the temperature decreases below the 
limit (approx. 75°C) 

Mains interruption (Pin 13) 

Mains has been interrupted (for 
the time longer than ~1 period) 

The fault signal will disappear after resetting 
INHIBIT signal 

Start-up (Pin 13) 

PCA has been just connected to 
the mains 

Once initial start-up 

sequence is performed 

completely

, i.e. in 2-5s, t

he signal is 

automatically removed

 

 

 

Содержание PCA-20

Страница 1: ...rning This equipment may be dangerous Please read the entire user manual carefully before using the product Important note Module is sensitive to the reverse polarity applied to the output If you aren...

Страница 2: ...ION 4 COOLING 4 APPEARANCE 4 CONNECTORS PINS INTERFACE SIGNALS 5 INSTALLATION 8 OPERATIONS 8 FAULTS 8 SPECIFICATIONS 9 DIMENSIONAL DRAWING 11 HOW TO ORDER 12 ENVIRONMENTAL PROTECTION 12 NAME AND ADDRE...

Страница 3: ...d to open the device Warning The equipment is not suitable for use in presence of a flammable anesthetic mixture with air or with oxygen or nitrous oxide Explanation of symbols Symbol Description Refe...

Страница 4: ...level VMAX could be choosen from 300V to 1500V at the moment of order Higher output voltages are available upon request By default modules are optimized for partial discharge applications Complete di...

Страница 5: ...Usage of longer cables may result in increased electromagnetic emission and or decreased electromagnetic immunity of the equipment and correspondingly in improper operation MOUNTING M4 thread x4 Modul...

Страница 6: ...imum output current 40mA 11 orange 5 V This pin provides 5V DC that may be used for status LEDs etc Maximum output current 40mA 12 18 N C 13 white black mark Overtemp Fault Indicator 2 This pin is pul...

Страница 7: ...5mA Rout 1kOhm 19 white red mark Fault Indicator This pin is pulled to ground when some failure occurs In this case the high voltage output will be also disabled Failures output short circuit overvol...

Страница 8: ...starts 4 Enable the high voltage output PIN7 of INTERFACE 5 After operations disable the high voltage output then disconnect module from the mains Faults Fault Cause of the fault Troubleshooting Overv...

Страница 9: ...ifications can be achieved in regime 70 100 of VMAX rated input voltage 25 C 2000W 2500W Nominal output power complete discharge modifications can be achieved in regime 10 100 of VMAX rated input volt...

Страница 10: ...urrent 200 A Isolations see Means of protections section Protections from open circuit from short circuit from overheating from overvoltage Cooling Forced air build in fan Environment Operation temper...

Страница 11: ...11 Dimensional drawing...

Страница 12: ...rge On demand PCA 20 800V PD 110 240VAC input 2000W 800V output partial discharge IEC 60601 1 2005 A1 2012 IEC 60601 1 2 2014 PCA 20 1000V PD 110 240VAC input 2000W 1000V output partial discharge IEC...

Страница 13: ...1000 4 3 List of frequencies Table 9 from 9 V m up to 28 V m PM 18 Hz or 217 Hz FM 1 kHz Complies Immunity from Electrical Fast transient EFT IEC 61000 4 4 2 kV on 230 VAC Tr Th 5 50 ns 100 kHz Compli...

Страница 14: ...king voltage depends on modification Input to input opposite polarity before fuses 1x MOOP 340Vpk Output to chassis Output negative is intentionally connected to the chassis Mains part to chassis 1x M...