EMC compliance
IEC 60601-1-2 Edition 4.0 (2014).
Environment of intended uses:
Professional Healthcare Facility Environment
Declared EMC compliance (to be confirmed with test reports for every particular
part number):
Test
Standard
Class/ Severity level
Test result
Emission
(IEC 60601-1-2 section 7.1-7.2)
Radiated emission
Freq. range: 30 - 1000 MHz
CISPR 11
Group 1 Class A
Complies
Conducted emission
Freq. range: 150 kHz - 30 MHz
Group 1 Class A
120 VAC / 230 VAC
Complies
Harmonic current emission
IEC 61000-3-2
230 VAC
Complies
Voltage changes, Voltage
fluctuations and Flicker test
IEC 61000-3-3
230 VAC
Complies
Immunity
(IEC 60601-1-2 section 8.9 & 8.10)
Immunity from Electrostatic
discharge (ESD)
IEC 61000-4-2
8 kV contact discharges &
15 kV air discharges
Complies
Immunity from radiated
electromagnetic fields
IEC 61000-4-3
3.0 V/m
80 MHz
2.7 GHz, 80% AM, 1kHz
Complies
Immunity from
Proximity field from wireless
communications equipment
IEC 61000-4-3
List of frequencies (Table 9),
from 9 V/m up to 28 V/m,
PM (18 Hz or 217 Hz), FM 1 kHz
Complies
Immunity from Electrical Fast
transient (EFT)
IEC 61000-4-4
± 2 kV on 230 VAC
Tr/Th – 5/50 ns, 100 kHz
Complies
Immunity from Surge
IEC 61000-4-5
±1.0 kV DM/ 2.0kV CM
on 230 VAC
Tr/Th – 1.2/50 (8/20)
s
Complies
Immunity from conducted
disturbances induced by RF fields
IEC 61000-4-6
3.0 & 6.0 V
RMS
on
230 VAC
0.15÷ 80 MHz, 80% AM, 1 kHz
Complies
Immunity from Voltage dips,
short interruptions and voltage
variations
IEC 61000-4-11
230 & 120 VAC mains;
0 % - 10 ms; 70% - 500 ms;
0% - 20 ms; 0% - 5sec
Complies
Note.
The emission characteristics of this equipment make it suitable for use in industrial
areas and hospitals (CISPR 11 class A). If it is used in a residential environment (for which
CISPR 11 class B is normally required) this equipment might not offer adequate protection to
radio-frequency communication services.
Warning.
Use of cables other than those specified or provided by the manufacturer of the
equipment may result in increased electromagnetic emission and/or decreased electromagnetic
immunity of the equipment and, correspondingly, in improper operation.