OEC UroView
®
2800 Table/Generator Functional Block Diagrams
78
Periodic Maintenance
MOTOR
(M2)
BRAKE
(NORMALLY APPLIED)
THERMAL SWITCH
(OPENS WITH HEAT)
FREQUENCY
INVERTER
(U1)
RELAY PCB
MOTOR ENABLE
SIGNAL FROM
CPU BOARD
K2
AC Input
X6
X9
X9
X2A
X2B
Q2
K1.2
22
21
R4
T2
(N.O. MOMENTARY)
K2
MASSE 24 V
X5
CPU PCB
S2
BRAKE ENABLE
1
2
3
3
16
16
5
3
1
13
14
2
4
6
A2-
A1+
LED 2
g
Page 1 of 1
06/02
GE OEC Training
Uroview 2800
Motor Drive Diagram
Tube Arm
10 K
1
1
Note:
S2 is the mechanical limit switch. It is a N.C.
switch which opens if the mechanical limit is reached.
During normal operation mechanical limits are never
reached. If the table has a problem and S2 is activated,
the T2 switch on the Relay PCB can be pressed
to bypass the limit switch and move the table off the
mechanical limit.
1
W
W
V
V
U
U
Three Phase
TUBARMTR.DS4
+24V
X8
LED 6
M24
FROM +G1
POWER
SUPPLY
F4
1
2
3
4
1
2
1
2
3 9
8
MASSE 24 V
Motor Drive Diagram – Tube Arm