Compact / Series 7600 Installation Procedure
9
X-ray Generator Information
The X-ray Generator in the Compact / Series 7600 complies with IEC 601-2-
7/1987 and CSA 22.2 #601.2.7. The following information provided in
accordance with these standards.
Reference: 601-2-7 Clause 6.8.2
1.
Maximum mA @ Nominal kVp
Standard Fluoro
3 mA @ 110 kVp
Dense Anatomy Fluoro
6 mA @ 110 kVp
Pulsed Fluoro
3 mA @ 110 kVp
Snapshot Fluoro
8 mA @ 110 kVp
Film Mode
20 mA @ 110 kVp
2.
Maximum kVp @ Maximum mA (See 1 above)
3.
Maximum Output Power
Snapshot (Fluorography)
0.88 kW @ 110 kVp, 8 mA
Film Mode
2.2 kW @ 110 kVp, 20 mA
4.
Nominal Output Power for 0.1 Second Exposure
Film Mode
2.0 kW @ 100 kVp, 20 mA
5.
Reference mAs @ 50% of Nominal Output Power
Film Mode
2.0 mAs @ 55 kVp
(20 mA, 0.1 second)
6.
(Not Applicable)
7.
(Not Applicable)
8.
Nominal Shortest Exposure Time
Film Mode
0.1 second