B64
IC25
+
-
kV
KVSOLL
6V
12V
Page 1 of 3
For Reference Only
kV GENERATION & REGULATION
3/7/97
KV.DS4
SWITCH CONTROL & DISPLAY LOGIC
COMPACT / SERIES 7600
DLIST
0kV
6V
36kV
7.98V
110kV
12.05V
55mV KVSOLL/kV
CAMERA
AU01 BOARD
Red
-kV
Yellow
+kV
16
CONTROL PANEL
+16V
LED10
R30
S10
kV
Manual
S11
kV
Increase
-KKVMAN
S12
kV
Decrease
-KVPLUS
-KVMINUS
ST1
11
12
13
B100
IC7
R28
ST2
ST1
2
6
KEYBOARD
INTERFACE
CONTROL
LOGIC
BUFFER IC4
PAL IC1
INVERTER IC11
KV
CONTOL
LOGIC
7
14
1
Relay
D7
14
7
Relay
D6
3
2
6
kV
DISPLAY
CONTROL
11
12
13
23A
22C
22A
X1
23A
22C
22A
16V
14A
-KKVMAN
-KVPLUS
-KVMINUS
Manual kV
Control
Auto kV
Control
17C
KVSOLL
17C
ST4
2
kV Control
to page 2
XPD
10
11
12
13
4 bit kV Display Code
IC4
H10
H11
H12
*100
*10
*1
kV DISPLAY
ST2
13
10
11
12
Relays shown in
de-energized condition
(Auto Fluoro Mode)
2
kV Calibration
During a 75kV manual fluoro exposure,
adjust R28 for 7.5 vdc on TPf (see page 2)
on the B54 board. Verify that there is 3.6 vdc
on TPf during a 36kV exposure and 10.0 vdc
during a 100kV exposure. You can measure kV
with the Keithly Triad System, using the wide
range filter pack or with the Victoreen 4000+.
2
KVSOLL Control Voltage
Adjustable by P1
KVSOLL
3
D4
U4-8
U11-13
K2-6
K1-6
U29-2
U24-5
KVQ1
KVQ2
KVQ3
KVQ0
KVQ1
KVQ2
KVQ3
KVQ0
Video Level signal from camera
typically 6vdc when ABS loop
is stablized.
(Compact / Series)
-XRAY
-XRAY
18C
K3-6
From Film Mode diagram
18C
D43
Xray-On (Green)
+12V
IC6
D10
+12V REG
P1
To page 3
KVSOLL
Adjust
From Fluoro Mode diagram
Used and Refurbished C-Arms are Available from www.SharpMedical.com - Call us at 800-969-9800