Page 1 of 3
For Reference Only
C-ARM LIFT
C-ARM UP MOVEMENT
3/7/97
CARMLIFT.DS4
COMPACT / SERIES 7600
S26
S29
UP SWITCH
CONTROL PANEL
ST1
38
-KCUP
BUFFER
A1
Y1
U4
+INCUP
11
PAL
U1
9
6
19
+CUP
B100 BOARD
D5
D9
T1
ST2
38
-CUP
ST7
5
D6
D4
D1
R4
C4
9
8
10
11
2
1
2
1
1
2
2
3
NC
NC
NO
UPPER LIMIT
SWITCH
(open at upper limit)
X3
5
X2
5
3
-CUP
B105 BOARD
NO
NO
NC
LIFT
MOTOR
L2
L4
L1
L3
6
7
23
24
14
13
NO
9
10
All relay contacts are shown in the proper
state for the lift motor to move up
T1
F4
10A
24VAC
BR1
AC Power In
POWER ON
Diagram
24VDC
X1
1
2
R6
R7
Gold resisters beneath the B105 board
1
1
K1
- Current Control
K2
- Direction Control Up
K3 - Direction Control Down
K4
- Current Enable Up
K5 - Current Enable Down
Energized
Energized
De-Energized
Energized
De-Energized
Used and Refurbished C-Arms are Available from www.SharpMedical.com - Call us at 800-969-9800