4c - Set PASS/FAIL Parameters
Use the Settings menu to set PASS/FAIL parameters.
0.00dB λ 1550nm
Touch “Settings”
Touch each dark rectangle to set a MIN and MAX for testing. MIN should be set to zero. MAX
allowable loss varies by build. Check your build documentation if you are not sure what to set for a
MAX loss. The most common MAX loss setting is 3dB. Touch OK when finished.
Any readings which fall between the MIN and MAX will be displayed in
GREEN
. Any readings which
are below the MIN or above the MAX will be displayed in
RED
. This is effectively a PASS/FAIL test.
0.50dB λ 1550nm
5.98dB λ 1550nm