11 Troubleshooting
16 / 24
Instruction Manual SDI® 4 (G/54320/EN 0220 Rev01)
Fault list - SDI® 4
Fault
Possible cause
Remedy
No function when the hand
switch / foot switch is pressed
(hand-held pushbutton for SDI® 4c
only optional)
The hand switch / foot switch
is not connected to the
SDI® 4e/c
no connection of the SDI® 4e/
c to the power supply
Power failure or power socket
is not active
These are not active if the
sockets of the stand are being
used.
Establish the connection to
the SDI® 4e/c
Establish the connection to
the power supply
Inform the in-house
electrician
Use the plug transformer
Switch the SDI® 4
mechanically
Activate the sockets in
accordance with the
instructions for the stand
Ask the microscope
manufacturer for assistance
The image is cropped
The SDI® 4 or other
components are incorrectly
mounted at an angle
The optics of the SDI® 4 are
not in operating position
(locking position)
Correct the assembly
Turn it on or off again with
the switch
Unclear image
The SDI® 4 or the optic of the
BIOM® 5
are dirty
Clean the glass surface of the
BIOM® 5 loupe
If the SDI® 4 optics are dirty,
contact OCULUS Service or
your authorized dealer.
The SDI® 4 or its fixtures wobble
The locking screws are loose
Tighten the locking screws
Fundus view is too narrow
Distance between
ophthalmoscopic lens and eye
too large
(only when using
BIOM® 5)
Magnification of the
microscope system to high
Carefully reduce the distance
using the microscope fine
adjustment mechanism
Reduce magnification of the
microscope
Pronounced reflections on BIOM®-
loupe or contact lens
The microscope light is on
Bright ceiling illumination
Turn the light off, illuminate
only intraocular