11 Measuring Procedure
44 / 85
Instruction Manual Pentacam® AXL (G/70100/EN 1121 Rev04)
Fine adjustment
Make any fine adjustments required based on the information in the
adjustment window. To do so, move the joystick in the specified
directions.
Example:
Move the joystick to the right.
Turn the joystick counter clockwise.
Move the joystick forwards.
When you have achieved the expected position, a black cross will
appear in the centre of the ring, surrounded by four black lines. The
Pentacam® will automatically begin measuring, alternately you can
start the measuring procedure manually.
For measuring manually: Start measuring by clicking [Start Scan] or
by pressing the return key.
Attention
It may not be possible to carry out a measurement with manual release.
Ask the patient to remove his or her head from the rest.
Check the measurement results by referring to the quality
specifications
.
Arrow Camera movement
Joystick movement
right
Move the joystick to the right
left
Move the joystick to the left
forward
Move the joystick toward the patient
back
Move the joystick away from the patient
up
Rotate the joystick clockwise
down
Rotate the joystick counter-clockwise