8 Set up and Connection
Instruction Manual Myopia Master® (G/68100/EN Rev04 0820)
17 / 92
8
Set up and Connection
Caution
Risk of incorrect measurements/equipment damage due to improper set-
up
Before first use, make sure the installation and connection of the
"Myopia Master®" examination station are completed by our service
team or by a professional authorized by OCULUS.
Note
Do not expose the Myopia Master® to any vibrations, shocks,
contaminants, moisture, or high temperatures.
Handle the optical device with care.
Set up the Myopia Master® so that the power plug is easy to access.
That way, you can easily disconnect it from the power supply for any
repairs or maintenance work.
Place the Myopia Master® on a level surface.
Place the device so that direct light cannot affect the measurement.
Make sure the examination is free from light reflections. To achieve
this, darken the examination room.
8.1
Electrical Connection
Caution
Electrical safety hazard
Do not use the Myopia Master® adjacent to or stacked with other
equipment.
If you have to use the Myopia Master® adjacent to or stacked with
other equipment, verify the correct operation of the Myopia
Master®.
Only use the power adapter listed in the list,
Only use a power cord which meets the requirements of IEC 60227-
1, type H05VVH2-F (type 53), minimum 0,75 m² and IEC 60320-1,
type C7.