5  Electromagnetic Compatibility (EMC) / Cables

Instruction Manual C-Quant ( G/80000/1007/en)

5 / 56

5

Electromagnetic Compatibility 
(EMC) / Cables

Special precautionary measures in respect of EMC apply to electrically operated medical
devices. 

No special measures need be observed in respect of OCULUS devices and systems.

Attention

Portable and mobile HF sets can interfere with electrically operated medical devices.

Cables

To minimise emitted interference and to increase interference resistance, the following
types of cables must be used:

6

Description of Unit and its 
Function

The C-Quant measures, in an accurate and objective way, the amount of straylight on the
retina caused by light scattering in a patient´s eye. In an ideal eye there would be no light
scattering at all, but because the eye media are not optically ideal, there will always be
some light scattering; which degrades the image projected on the retina, thus the quality
of vision. 

The effect of increased straylight can be compared to what would be seen through very
dirty glasses. In short, more straylight means worse vision.

Straylight may be significantly increased above normal values because of cataract, 
corneal distrophies, refractive surgery or other pathologies.

It is designed for computer-controlled operation via serial or USB interface.

7

Operation

7.1

Before first Use

Please remove the device as well as the accessories from the packaging and keep this. You
can use it for correctly sending or transporting the device in the case of service or repair.
In this way you avoid unnecessary damage and costs. 

Read these operating instructions carefully. 

05200320 or 05200210 (110 volt)

Mains cable

2.5 m in length

Содержание C-QUANT

Страница 1: ...OCULUS C QUANT INSTRUCTION MANUAL...

Страница 2: ...ed in cooperation with Dr Tom van den Berg PhD from the Netherland Ophthalmic Research Institute of the Royal Academy NORI in 2005 merged into the netherlands institute for Neuroscience www nin nl in...

Страница 3: ...a Management 7 8 1 Starting Patient Data Management 7 8 2 Entering new patients 7 8 3 Selecting existing patients 8 8 4 Deleting examinations 9 8 5 Patient data 9 8 6 Exporting patient data 10 8 7 Imp...

Страница 4: ...r operating the C Quant 35 12 1 Measuring procedure 35 12 1 1 Patient instruction 36 12 1 2 After the measurement 38 12 2 Measurement Examples 40 13 Troubleshooting 49 14 Inserting corrective lenses 4...

Страница 5: ...t This chapter contains a summary of the most important points regarding these safety matters Attention Do not use the device if you have not understood the instruction manual Make sure to keep this i...

Страница 6: ...rvice department 2 2 Pictogram definitions Attention Identifies a potentially dangerous situation which may cause minor injury or damage to property Note Instructions for use and useful or important i...

Страница 7: ...rate the C Quant in rooms used for medical purposes if VDE 0100 installa tion procedures have been observed Do not force any plug connections If you are unable to make a plug connection check whether...

Страница 8: ...are to observe the safety precautions given above 4 Transportation and Disposal 4 1 Transport and storage instructions Great care must be taken when transporting the C Quant from one location to anoth...

Страница 9: ...be no light scattering at all but because the eye media are not optically ideal there will always be some light scattering which degrades the image projected on the retina thus the quality of vision...

Страница 10: ...nt with the PC via the serial or the USB cable Connect the low voltage socket of the supplied desktop power supply to DC input connector Connect the AC power cable to the desktop power supply Connect...

Страница 11: ...en displays the user interface for Patient Data Management In case the user interface does not start automatically Click the C Quant icon on the desktop Fig 8 1 Patient Data Management Note To enter t...

Страница 12: ...alphabetical order Click Search to quickly find the patient you require from the list Enter the patient s name or the first letter of the name in the Last name field Optionally you can search for the...

Страница 13: ...e to delete individual examinations from the patient data Click Delete exam underneath the list of examinations You will be asked if you really want to delete the examination Fig 8 4 Delete individual...

Страница 14: ...asurement data to another storage medium e g USB drive Sending stored data by e mail In Patient Data Management select the patient you require You can select a patient s entire data master or individu...

Страница 15: ...e program is updated on the C Quant PC sender the program on the other PC recipient must also be updated 8 7 Importing patient data Import the patient and examination data you have received into your...

Страница 16: ...F for a network drive Saving data can be done in a number of ways Save all data If this option is selected all examination and patient data are saved Changed and new data only If this option is selec...

Страница 17: ...f the source from which the data are to be restored Click on the button to select the drive and directory from which the data are to be restored Click the Restore button for uploading the data from th...

Страница 18: ...ngs are made here Fig 8 8 Main Setting Language Specify the language which the program should use Optional items for patient Optional information can be entered here Such information can then be activ...

Страница 19: ...can be set up for different configurations The settings in the following fields can then be set separately for each configuration The user is then selected from the drop down menu and is activated by...

Страница 20: ...nt data is always created locally on your PC and is used as a transfer folder for transfer of the data to the database Here again the name is always the name of the device followed by the extension DA...

Страница 21: ...Alternatively the drive and the directory from which the data are to be imported or exported can be searched for by pressing the button Folder in field Import Export This field is for entering the let...

Страница 22: ...nsfer takes place via the standard E mail function of your computer SMPT If an E Mail program is not installed on your PC activate the option SMTP The data transfer then takes place directly based on...

Страница 23: ...80000 1007 en 19 56 8 9 5 Settings Interface Fig 8 12 Interface Box On this tab panel you can specify the storage locations for various interface types These interfaces are intended for the integrati...

Страница 24: ...d reader you can read your patient data directly from the patient s insurance card into the OCULUS Patient Data Management Click on the Enable checkbox To install and configure your smart card reader...

Страница 25: ...your network does not meet the system requirements to run Version 6 x then the use of both versions in the network is unavoidable In that case just one PC with Version 6 x must be configured such tha...

Страница 26: ...start the Examination program by clicking C Quant Alternately you can start the program by double clicking the patient you have selected 9 C Quant Program Fig 9 1 C Quant program menu The menu bar con...

Страница 27: ...left corner of the main window Fig 9 2 Common data 9 1 2 Examination specific data Fig 9 3 Relevant data to the examination This is where to enter all data relevant to the examination Eye Select the...

Страница 28: ...the patient responds otherwise the program issues a warning The third test stimulus is difficulty for many patients Click Continue anyway if the patient shows proper understanding Fig 9 5 Warning mis...

Страница 29: ...hows the examination date the patient s age at the time of exam ination the examination duration and the quality parameters Esd and Q Esd is the expected standard deviation of the individual measureme...

Страница 30: ...the gray area gives the range of normal variation If the red dot is outside the gray area this signifies a conspicuous finding 9 1 4 Examination fields Stimulus field view Fig 9 8 Stimulus field This...

Страница 31: ...gram are represented by different symbols depending on the phase currently being examined instruction phase dark phase light phase Extensive studies have shown that the results follow a specific curve...

Страница 32: ...e is projected onto the retina However some of the light is scattered by the lens and other parts of the eye and is thus projected on the part of the retina onto which the test field projects fovea On...

Страница 33: ...se Now the brightness of the test field is adjusted until the patient no longer sees any flicker This yields his or her straylight value 10 2 Compensation comparison method Determining the exact stray...

Страница 34: ...is disregarded in this example because it has no significance for understanding the measurement principle Seven stimuli for the compensated field are shown Fig 10 2 Seven stimuli The left test field c...

Страница 35: ...as intense as the scattered light meaning that the flicker intensity is 20 10 10 equal to that in the other test field The patient will therefore find that the two test fields flicker with equal inten...

Страница 36: ...onfusing differences in flicker intensity with differenc es in brightness the instrument continually adjusts the test fields so that they have equal average brightness at each presentation The absciss...

Страница 37: ...esult of the dark phase gives a reasonably good approximation of the patient s stray light value To determine it with greater precision he is now shown 13 stimuli in the light phase which are grouped...

Страница 38: ...usly adjusted When the examination has been completed the result fields show the final values If during the examination the patient fails to respond to a stimulus within two seconds the following warn...

Страница 39: ...his chapter gives some examples of possible measurement outcomes which can be use ful to interpret your own results especially when measurement reliability is not optimal It is assumed that you are al...

Страница 40: ...y cases if straylight increase is mild If high straylight values are expected e g with cataract or corneal turbidities settings F medium log s or G high log s must be used G can also be used if you wa...

Страница 41: ...nd also here you must make a choice following your first impulse Optional information only if the patients ask one measurement consists of about 25 presentations and usually takes 1 5 to 2 minutes Mak...

Страница 42: ...0 08 Before repeating the measurement you should verify if a correct Range setting was chosen see below If esd 0 12 the measurement should be considered not reliable but see below The measurement sho...

Страница 43: ...ting the measurement with the same Range setting will not yield a lower esd value You can only improve reliability by choosing a different Range setting Note It is not always clear if the chosen range...

Страница 44: ...40 56 Instruction Manual C Quant G 80000 1007 en 12 2 Measurement Examples Example 1 esd 08 and Q 1 This is a reliable measurement Note that there is almost no overlap between the 0 and 1 responses Mo...

Страница 45: ...le 2 0 08 esd 0 12 Although reliability is not optimal this measurement may be accepted as a good measurement in most cases However if time allows it is recommended to repeat the measurement in order...

Страница 46: ...easurement this measurement can be used to es timate an upper limit for the straylight value Because there is a consistent row of 1 responses in the right part of the graph it is quite certain that th...

Страница 47: ...e of this measurement can be used to estimate an upper limit for the straylight value If you decide to repeat the measurement the surplus of 1 responses would suggest choosing a lower Range setting as...

Страница 48: ...xamples 3a and 3b only esd is so high that no estimation for the straylight value could be calculated log s 6 00 is a meaningless value However you can make this estimation yourself by looking at the...

Страница 49: ...00 1007 en 45 56 Example 4 Range setting too high This measurement went very well only there are too many 1 responses and not enough 0 responses for a reliable estimation of the straylight value Relia...

Страница 50: ...t G 80000 1007 en Example 5 Range setting too low There are too many 0 and not enough 1 responses for a reliable estimation of the straylight value The measurement reliability can be improved by repea...

Страница 51: ...Example 6 unreliable measurement at low Range setting From this result you can not tell if the Range setting was chosen too low or if the test is too difficult for the patient In both cases it is wise...

Страница 52: ...uction Manual C Quant G 80000 1007 en Example 7 reliable measurement at high Range setting The same patient as in Example 6 but measured with the G setting This is a reliable measurement esd 0 08 and...

Страница 53: ...No function after the mains switch is turned on the mains switch does not light up C Quant is not connected to the mains supply There has been a power failure or the power outlet is not active Plug t...

Страница 54: ...n use Cleaning enameled surfaces Take care not to let cleansers enter the instrument We generally recommend using an antistatic cleanser for plastic surfaces to inhibit recontamination Otherwise you c...

Страница 55: ...oft cloth 16 Terms of Warranty and Servicing 16 1 Terms of warranty The C Quant you have purchased is a high quality OCULUS product This C Quant was carefully manufactured using quality materials and...

Страница 56: ...ly extensions adjustments changes or repairs are carried out by unauthorised personnel if the C Quant is maintained improp erly or if it is handled incorrectly If the work described above is carried o...

Страница 57: ...ion of Conformity We declare in our responsibility that this product complies with the following norms or normative documents EN 60601 1 EN 60601 1 2 as stipulated by Guideline No 93 42 EEC for medica...

Страница 58: ...g Dimensions W x D x H 250 x 325 x 380 mm Interface RS 232 V 24 sub D USB 1 1 Typ B Type of protection against electrical shock Protection class 1 Level of protection against electrical shock Type B L...

Страница 59: ...L van Rijn LJ van den Berg TJTP Reliability of the compensa tion comparison stray light measurement method J Biomed Opt 11 34027 2006 Van den Berg TJTP van Rijn LJ Michael R Heine C Coeckelbergh T Ni...