2 Safety Instructions
Conditioning Instruction BIOM® 5 (G/55185/0818/en)
3 / 26
Conditioning Information
Attention
Risk of injury if the BIOM® 5 is not sterile
If the patient or his bodily fluids come into contact with the BIOM® 5, it
can become contaminated, e.g. when putting it away.
Make sure that the BIOM® 5 is cleaned, disinfected and sterilised.
Condition the BIOM® 5 before the first and every subsequent use.
Comply with the legal provisions in force in your country, and with
the hygiene and waste disposal regulations of the hospital or clinic.
Condition the BIOM® 5 only after you have fully understood this
instruction manual.
The BIOM® 5 and all sterilisable components of the BIOM® 5 must
be cleaned, disinfected and sterilised prior to initial use and prior to
every subsequent use. To do so, take the BIOM® 5 out of the
packaging.
Make sure that only validated device and product-specific
procedures are used for cleaning/disinfection and sterilisation and
that the validated parameters are observed for each cycle.
Use a machine (disinfector) for cleaning/disinfection purposes. This
is much more effective.
Notes on returning
Attention
Personal injury caused by contaminated BIOM® and components
Before returning the product to OCULUS: Prepare the BIOM® and
sterilisable components according to this treatment instructions.
Send only visibly prepared OCULUS products back to OCULUS