3 Safety Instructions
Instruction Manual SDI® 4 (G/54320/EN 0220 Rev01)
5 / 24
Operation and Maintenance Information
Before initial operation: Let OCULUS or an authorized dealer train
you in the operation of the SDI® 4.
Never operate a damaged SDI® 4. Inform the supplier of any damage.
Only operate the SDI® 4 with the original accessories supplied by us
and only when the unit is in a technically perfect condition.
Only operate the equipment after your have read and understood the
Instruction Manual.
Do not touch the patient and the device at the same time.
If a malfunction occurs that you cannot rectify by following in the
instructions in the troubleshooting table
, you
must not use the device. Clearly mark the unit as non-operational
and get in touch with our service personnel.
The sterilisable rubber cap for the adjusting knob of the SDI® 4 must
be sterilised before the first and before each subsequent use.
It is imperative that you heed the cleaning, disinfection and
sterilisation instructions given in the Conditioning Manual.
A sterilisation of the SDI® 4 is not possible.
Instructions for disassembly and disposal
When disconnecting electrical connections, pull on the respective
plug and not on the cable itself. Take hold of the plug, do not pull on
the cable.
Dispose of the equipment in compliance with the corresponding
legal requirements. Comply with the hygiene and disposal
regulations of the hospital or clinic.
Patient environment information
Patient environment is the area where patients can come into contact
with any part of a medical electrical equipment (ME equipment) or with
another person being in contact with the ME equipment.
In the patient environment, use devices that conform to IEC 60601-1. If
a multiple power socket is to be used, or if a device is to be used that
does not meet the IEC 60601-1 standard, use an isolating transformer.