E101V USER MANUAL
15
Version 00.30 7/4/2016
6.5
HRU Replacement
If you have an EPIRB mounted in a float free housing, this will also contain a HR1E
Hydrostatic Release Unit (HRU). The HRU unit must be replaced two years after installation
- the expiry date is marked on the HRU and on the front of the housing. If this date has
been reached then the HRU must be replaced with an Ocean Signal HR1E, failure to do so
may result in the HRU not operating correctly during an emergency situation.
1.
Lift the release mechanism by
pulling against the spring and
remove the EPIRB from the housing.
2.
Push the HRU down against the
spring and remove the locking pin.
Carefully remove the HRU from the
spring.
3.
Using the new HRU, locate the two
retaining ridges (at the bottom of the
HRU) onto the spring. Carefully push
the HRU against the spring.
4.
Push the HRU into position as shown.
Push the locking pin home with the
retaining flange pointing down. Load
the EPIRB into the housing.
Mark the label on the HRU and the supplied label for fitting to the ARH101, with the expiry
date using a UV resistant indelible pen. The expiry date should be two years after the date
of installation, but no more than three years after the date of manufacture marked on the
HRU.