21
MAINTENANCE & SERVICE
MAINTENANCE
& SERVICE
WIRING DIAGRAM
ELECTRICAL SPECIFICATIONS
Designed for:
North America
Europe
Input Service
120 VAC/15 amp/60Hz
220 VAC/10 amp/50/60Hz
Current Draw
4.0 amps
2.0 amps
Maximum
Momentary Current
Consumption
6.0 amps
3.0 amps
Voltage Output to
Actuators
24 VDC
24 VDC
Electric Shock
Protection
Class 1 Equipment
Class 1 Equipment
Tabletop Applied Part
Type B Applied Part
Type B Applied Part
Table Top IEC 529 Rating
IPXO
IPXO
Mode of Operation
Intermittent Operation
MAX 2 minutes ON
18 minutes OFF
Intermittent Operation
MAX 2 minutes ON
18 minutes OFF
PERMISSIBLE ENVIRONMENTAL CONDITIONS
Conditions
Temperature
Humidity
Atmospheric Pressure
Normal Use
50° (10°C) to 104° (40°C)
20% to 60% RH
98 to 105 kPa
Storage & Transport
-20° (-29°C) to 135° (57°C)
20% to 95% RH
98 to 105 kPa
TABLE SPECIFICATIONS
North America
Europe
Weight
~ 235 lbs.
~ 107 kg
Shipping Weight
~ 325 lbs.
~148 kg
Lifting Capacity
550 lbs.
250 kg.