OAKWORKS DTPM300 Скачать руководство пользователя страница 13

11

rePAir ProceDUreS

5.  Reassemble in reverse order. 

6.  Plug table into power and test all operations of the table.

7.  Tuck and secure extra cable wire up out of the way. Make sure all cables are not in a pinch area.

8.  It may be necessary to reset the control box. See table resetting:

a. 

TABLE RESET: TABLES MANUFACTURED BEFORE 03/15/12

b. 

TABLE RESET: TABLES MANUFACTURED 03/15/12-02/25/13

c. 

TABLE RESET: TABLES MANUFACTURED AFTER 02/25/13

9.  Please fill out and return a 

Service Replacement Form

.

table of Contents

Содержание DTPM300

Страница 1: ...Service MANUAL DTPM300 Tables Part No MMMNSM0002 Revision Level F...

Страница 2: ...11 Hand Control Foot Control Replacement 12 13 Lift Column Replacement 14 16 Caster Replacement 17 19 Schematic 20 Wiring Diagram 21 Replacement Part s List 22 Revision History back cover Contact Info...

Страница 3: ...ZZ Ser FLR561910 DTPM RT 24 x 24 40 x 78 MG TT TTCoal 01 04 2013 Order 654321 Notes Sample Product Label Model Numbers and Serial Numbers always start with a letter The model number and serial number...

Страница 4: ...epeat using the foot control Check for abnormal noises Replace any missing or illegible labels Check that all fasteners are present and fastened securely Check table grounding Clean unusual buildup of...

Страница 5: ...le manufactured after 02 25 13 3 See Control Box Replacement If the hand control and or foot control are not working 1 Defective Hand Foot Control 1 See Hand Foot Control Replacement If one or both of...

Страница 6: ...below 4 Check all connections on the control box of the table See photo below 5 Make sure that all male and female connectors are firmly pushed into each other Even a slight misconnection can cause po...

Страница 7: ...e buttons when the beeping stops 2nd row buttons Down button 2 Run all the motors in the system to the fully retracted position Down Caution Both columns should move down unless already fully down If...

Страница 8: ...imes it takes sev eral attempts to press them simultaneously The control box will beep approximately 19 times You must press buttons to level or retract columns within 10 seconds or it will exit manua...

Страница 9: ...ks for further assistance if your table does not reset 717 235 6807 Down button 3 Run all the motors in the system to the fully retracted position Down Caution If the tilt on the table starts to Incre...

Страница 10: ...ximately 10 times Release the buttons when the beeping stops Top row buttons Down button 2 Run all the motors in the system to the fully retracted position Down Caution If the tilt on the table starts...

Страница 11: ...source Lastly give yourself adequate room around the table to make the task more comfortable and safe PLEASE READ INSTRUCTIONS COMPLETELY BEFORE REPAIR 1 Identify and mark cables Masking tape works g...

Страница 12: ...this to insure control box has power Photo 1 Activated Photo 2 Depressed 3 Remove the strain relief and all cable from the control box Remove ground wire from control box and disconnect power cord usi...

Страница 13: ...ut of the way Make sure all cables are not in a pinch area 8 It may be necessary to reset the control box See table resetting a TABLE RESET TABLES MANUFACTURED BEFORE 03 15 12 b TABLE RESET TABLES MAN...

Страница 14: ...CTIONS COMPLETELY BEFORE REPAIR 1 Hand Control Replacement Remove Strain Relief Plug the new Hand Control into the Control Box Test new Hand Control and if this does not work see Control Box Replaceme...

Страница 15: ...control box See table resetting a TABLE RESET TABLES MANUFACTURED BEFORE 03 15 12 b TABLE RESET TABLES MANUFACTURED 03 15 12 02 25 13 c TABLE RESET TABLES MANUFACTURED AFTER 02 25 13 4 Please fill out...

Страница 16: ...stly give yourself adequate room around the table to make the task more comfortable and safe PLEASE READ INSTRUCTIONS COMPLETELY BEFORE REPAIR It will be necessary to power the table for varies tasks...

Страница 17: ...functioning column end of the table 3 2 people required Carefully roll the top and lift off of the main bearings then roll the rest of the way till the top lifts off completely The top is heavy and aw...

Страница 18: ...thout lifting the base from the floor and above the base for the strain relief and power cord removal 6 Reassemble in the reverse order Torque bolts in column bottom and top to 15 Ft lbs 7 Slide the t...

Страница 19: ...ure that the table is clear of all people and materials before attempting to repair Remove the power cable from its power source Lastly give your self adequate room around the table to make the task m...

Страница 20: ...lding the caster to keep it from rotating 4 Raise the base and insert a wood spacer underneath so the caster is not touching the ground 5 Unscrew the locked caster It should break free easily with han...

Страница 21: ...ighten with a socket wrench turn past flush 8 Using a general purpose cement glue down the black rubber plastic cover We use Loctite super glue gel Do not get gel on the screw threads 9 Please fill ou...

Страница 22: ...EDGES UNLESS OTHERWISE NOTED ALL DIMENSIONS ARE IN INCHES X X 03 X XX 02 X XXX 005 ANGLE 1 DO NOT SCALE DRAWING AUTHOR OAKWORKS INC NOTES DESCRIPTION REVISIONS REV DATE MATERIAL DTPM 300 SCHEMATIC DTP...

Страница 23: ...21 Wiring diagram Table of Contents...

Страница 24: ...Control box 230V Systems 1 57916 2 Control Box CB16 220V Openbus Hand Control 1 Each 57920 57920 1 Hand Control 4 Button Hand Control Overlay 4 Button DTPM Foot Control 1 4027 Foot Control 2 Button L...

Страница 25: ...tion IEC STD 60601 1 6 3rd Edition Certified to CAN CSA STD C22 2 No 60601 1 CB TEST CERTIFICATE AND REPORT Service MANUAL dtpm300 Tables Part No MMMNSM0002 Revision Level F Revision History Revision...