SPECIFICATIONS

Guidance and manufacturer’s declaration - electromagnetic immunity

The table is intended for use in the electromagnetic environment specified below. The customer or the user of the 

table should assure that it is used in such an environment.

Immunity Test

IEC 60601 test level

Compliance level

Electromagnetic

Environment - guidance

Conducted RF 

IEC 61000-4-6

Radiated RF

IEC 61000-4-3

3 Vrms

150 kHz to 80 MHz

3 V/m

80 MHz to 2,5 GHz

3Vrms

3V/m

Portable and mobile RF communications equipment 

should be used no closer to any part of the table, includ-

ing cables, than the recommended separation distance 

calculated from the equation applicable to the frequency 

of the transmitter.

Recommended separation distance

d = 1.2 

 P

 

d = 1.2 

 P

   80 MHz to 800 MHz

d = 2.3 

 P 

  800 MHz to 2,5 GHz

Where P is the maximum output power rating of the trans-

mitter in watts (W) according to the transmitter manufac-

turer and d is the recommended separation distance in 

meters (m).

Field strengths from fixed RF transmitters, as deter-mined 

by an electromagnetic site survey, 

a

 should be less than 

the compliance level in each frequency range. 

b

Interference may occur in the vicinity of equipment 

marked with the following symbol:

NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies.

NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection 

from structures, objects and people.

a

  Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios, 

amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To assess the 

electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be considered. If the measured 

field strength in the location in which the table is used exceeds the applicable RF compliance level above, the table should be 

observed to verify normal operation. If abnormal performance is observed, additional measures may be necessary, such as re-

orienting or relocating the table

b

  Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 1 V/m.

30

Содержание OAKWORKS ProLuxe Series

Страница 1: ...USER MANUAL www oakworks com 717 235 6807 OAKWORKS ProLuxe Series...

Страница 2: ...within this document is subject to change without notice and should not be construed as a commitment by OAKWORKS Inc OAKWORKS Inc encourages requests for technical specifica tions and the like documen...

Страница 3: ...Cabi Storage Shelf 15 Removing Top Pad 16 Adjustable Breast Comfort ABC System Option 17 Integrated Heated Top Option 18 QuickLock Face Rest Accessory Option 19 Adjustable Side Arm Rests Accessory Op...

Страница 4: ...ON IMPORTANT SAFETY INSTRUCTIONS PRODUCT USE DESCRIPTION Oakworks ProLuxe Series offers a variety of positioning capabilities for massage therapists and spa professionals of all specialties On these m...

Страница 5: ...not use attachments not rec ommended by the manufacturer 7 To disconnect turn all controls to the off position then remove plug from outlet 8 Improper connection of the equipment grounding conductor...

Страница 6: ...or should read these sections carefully Weight Limit client and accessories 550 lbs 250 kg Do not exceed Adjustable Breast Comfort ABC System Weight Limit 100 lbs 45 kg Do not sit on the ABC System On...

Страница 7: ...like the plug shown below See Grounding Methods US 120V Example Make sure that the product is connected to an outlet having the same configuration as the plug No adapter should be used with this produ...

Страница 8: ...ble Capacity 550 lbs 250 kg client and accessories Widths 29 74 cm 31 79 cm or 33 84 cm Length 73 185 cm Padding 4 10 cm or 6 15 cm Aerocel Foam Table Weight 200 lbs 91 kg Height Range with Cabi Stora...

Страница 9: ...ease Operate the foot and or optional hand control as shown below to raise or lower the height backrest footrest or optional ABC functions of the table Backrest Down Backrest Up Height Down Height Up...

Страница 10: ...Capacity 550 lbs 250 kg client and accessories Widths 29 74 cm 31 79 cm or 33 84 cm Length 73 185 cm Padding 4 10 cm or 6 15 cm Aerocel Foam Table Weight 190 lbs 86 kg Height Range with Cabi Storage...

Страница 11: ...1 P N 71569 1 P N 71568 1 Height Down ABC Down ABC UP P N 71567 1 P N 71569 1 P N 71568 1 P N 71567 2 P N 71569 2 ABC DOWN P N 71567 1 P N 71569 1 P N 71568 1 P N 71567 2 P N ABC SYSTEM UP ABC SYSTEM...

Страница 12: ...ty 550 lbs 250 kg client and accessories Widths 29 74 cm 31 79 cm or 33 84 cm Length 73 185 cm Padding 4 10 cm or 6 15 cm Aerocel Foam Table Weight 180 lbs 82 kg Height Range with Cabi Storage Shelves...

Страница 13: ...P N 71567 1 P N 71569 1 P N 71568 1 Height Down ABC Down ABC UP P N 71567 1 P N 71569 1 P N 71568 1 P N 71567 2 P N 71569 2 ABC DOWN P N 71567 1 P N 71569 1 P N 71568 1 P N 71567 2 P N ABC SYSTEM UP...

Страница 14: ...Hz 230V 50Hz Table Capacity 550 lbs 250 kg client and accessories Widths 29 74 cm 31 79 cm or 33 84 cm Length 73 185 cm Padding 4 10 cm or 6 15 cm Aerocel Foam Table Weight 170 lbs 70 kg Height Range...

Страница 15: ...ate the optional ABC DOWN UP Height UP P N 71569 1 P N 71568 1 Height DOWN P N 71567 1 P N 71569 1 P N 71568 1 Height Down ABC Down ABC UP P N 71567 1 P N 71569 1 P N 71568 1 P N 71567 2 P N 71569 2 A...

Страница 16: ...that clients take some of their weight of the leg rest while you lift TO LOWER THE LEG REST Pull the lever on the side of the leg section and push the leg section down Release the lever once the leg r...

Страница 17: ...le top to protect the surface from any debris from the bottom of your foot control Two people are required to safely move the table Grasp the table frame securely at the end with the molded rubber fee...

Страница 18: ...is located and the head end of the table The Cabi Shelf holds a Hot Towel Cabi with dimensions no greater than 17 9 W x 12 6 D x 11 4 H Place the Hot Towel Cabi on the Cabi Shelf and center it Wrap t...

Страница 19: ...crews If the thumbs screws are on too tight a 3 16 bit can be used 2 Lift top pad off the platform IF THE TABLE HAS A HEATED TOP Unplug the table from the wall outlet and disconnect the wires circled...

Страница 20: ...ometimes going back up a little will be preferred This whole process takes no more than 15 20 seconds After lowering check headrest position to ensurecomfort support NOTE Breastsizeisnottheonlydetermi...

Страница 21: ...f the pad while Zone 2 controls the lower half of the pad Press both but tons located on the right side of the table in the seat section to turn on both zones and heat the entire top BLINKING LIGHTS I...

Страница 22: ...to the bracket until bullet catches are snapped into place To remove the QuickLock Face Rest gently pull the headrest by the dowels until it comes free Set it safely aside or store it inside the under...

Страница 23: ...rest at a 60 angle to the backrest 2 Push end all the way into the hole and drop the arm rest into the horizontal position Pull outward to make sure the arm rest is secure TO LEVEL THE ARM RESTS The...

Страница 24: ...head support Note The QuickLock Face Rest is required to use the boiance shelf support system TO INSTALL THE BOIANCE SHELF SUPPORT First install the QuickLock Face Rest at the head end of the table 1...

Страница 25: ...d end of the table Position Bluetooth Speaker with the magnetic attachment facing up Plug the cord into the side of the speaker and attach it to the bracket CONNECTING TO BLUETOOTH SPEAKER Turn Blueto...

Страница 26: ...edure Because of possible chemical incompatibilities between various cleaning products and bleach utmost care must be taken by the user to avoid potential exposure to harmful or toxic by prod ucts of...

Страница 27: ...o wipe off any excess grease RECOMMENDED PERIODIC INSPECTIONS yearly or local standard Check for damage to the power hands control and foot control cables and all visible wiring Visually inspect compo...

Страница 28: ...number are located on the underside of the foot section Serial Number 16796025 PXXA4EX1T410USXXXXXX Ser PXR563146 Proluxe Electric Salon w ABC 1 T 4 10 US XX XX XX TTSKYBLUE 01 01 2020 Order 654321 No...

Страница 29: ...LE DIMENSION ELECTRIC SALON LIFT ASSIST 26 TABLE DIMENSION FLAT TOP CABI SHELF DIMENSIONS 19 5 L x 9 W x 11 88 H 49 5cm L x 23cm W x 30 25cm H STORAGE SHELF DIMENSIONS 19 5 L x 13 W x 8 63 H 49 5cm L...

Страница 30: ...57 C 20 to 95 RH 98 to 105 kPa ELECTRICAL SPECIFICATIONS Designed for North America Europe Input Service 120 VAC 15 amp 60 Hz 230 VAC 10 amp 50 60 Hz Current Draw 2 7 amps 1 5 amps Maximum Momentary...

Страница 31: ...ectromagnetic environment in which radiated RF disturbances are con trolled The customer or the user of the table can help prevent electromagnetic interference by maintaining a minimum distance betwee...

Страница 32: ...al mode 2 kV common mode 1 kV differential mode 2 kV common mode Mains power quality should be that of a typical commercial or hospital environment Voltage dips short interrup tions and voltage variat...

Страница 33: ...F transmitters as deter mined by an electromagnetic site survey a should be less than the compliance level in each frequency range b Interference may occur in the vicinity of equipment marked with the...

Страница 34: ...31 THIS PAGE IS INTENTIONALLY LEFT BLANK...

Страница 35: ...32 THIS PAGE IS INTENTIONALLY LEFT BLANK...

Страница 36: ...MS TO ANSI AAMI STD ES60601 1 IEC STD 60601 1 3RD EDITION IEC STD 60601 1 2 3RD EDITION IEC STD 60601 1 6 3RD EDITION CERTIFIED TO CAN CSA STD C22 2 NO 60601 1 CB TEST CERTIFICATE AND REPORT OAKWORKS...

Отзывы: