SPECIFICATIONS

Guidance and manufacturer’s declaration - electromagnetic immunity

The table is intended for use in the electromagnetic environment specified below. The customer or the 

user of the table should assure that it is used in such an environment.

Immunity Test

IEC 60601 test level

Compliance level

Electromagnetic

environment - guidance

Conducted RF 

IEC 61000-4-6

Radiated RF

IEC 61000-4-3

3 Vrms

150 kHz to 80 MHz

3 V/m

80 MHz to 2,5 GHz

3Vrms

3V/m

Portable and mobile RF communications equipment 

should be used no closer to any part of the table, in-

cluding cables, than the recommended separation 

distance calculated from the equation applicable 

to the frequency of the transmitter.

Recommended separation distance

d = 1.2 √ P

 

d = 1.2 √ P

   80 MHz to 800 MHz

d = 2.3 √ P 

  800 MHz to 2,5 GHz

Where P is the maximum output power rating of the 

transmitter in watts (W) according to the transmitter 

manufacturer and d is the recommended separa-

tion distance in meters (m).
Field strengths from fixed RF transmitters, as deter-

mined by an electromagnetic site survey, 

a

should be 

less than the compliance level in each frequency 

range. 

b

Interference may occur in the vicinity of equipment 

marked with the following symbol:

NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and 

reflection from structures, objects and people.

a

  Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile 

radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accura-

cy. To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be 

considered. If the measured field strength in the location in which the table is used exceeds the applicable RF com-

pliance level above, the table should be observed to verify normal operation. If abnormal performance is observed, 

additional measures may be necessary, such as re-orienting or relocating the table

b

  Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.

Содержание PF Series

Страница 1: ...L www oakworksmed com 717 235 6807 made in the USA with US imported parts PF250 PF200 Shown with optional QuickLock Face rest platform and Boiance crescent PF400 Shown with optional Adjustable Side Ar...

Страница 2: ...within this document is subject to change without notice and should not be construed as a commitment by OAKWORKS Inc OAKWORKS Inc encourages requests for technical specifications and the like documen...

Страница 3: ...rest Adjustment 12 Lowered Arm Rest Use 12 Locking Casters Use 13 Table Extender 13 Adjustable Arm Rest Installation and Adjustment 14 QuickLock Face Rest 15 Cleaning Disinfection Recommended Cleaners...

Страница 4: ...Terminal This symbol when used in this manual or on product labels indicates that the product should be protected from moisture The humidity specifications for Transport Storage are listed in the env...

Страница 5: ...thetic mixture with air or oxygen or nitrous oxide Do not lift the table by the table top This can damage the table When lowering the table or using the Trendelenburg functions make sure there is noth...

Страница 6: ...0 kg patient and accessories Widths 27 69 cm 29 74 cm 31 79 cm or 33 84 cm Length 4 section 72 183 cm Padding 4 10 cm Comfort Foam Upholstery Terra Touch Wood Stain Natural Maple Finish Table Weight 2...

Страница 7: ...s 250 kg patient and accessories Widths 27 69 cm 29 74 cm 31 79 cm or 33 84 cm Length 2 section 74 188 cm Padding 3 8 cm Comfort Foam Upholstery Terra Touch Wood Stain Natural Maple Finish Table Weigh...

Страница 8: ...patient and accessories Widths 27 69 cm 29 74 cm 31 79 cm or 33 84 cm Length 2 section 73 185 cm Padding 3 8 cm Comfort Foam Upholstery Terra Touch Wood Stain Natural Maple Finish Table Weight 212 lb...

Страница 9: ...is equipped with a cord having an equipment grounding conductor and a grounding pin The pin must be plugged into an appropriate outlet that is prop erly installed and grounded in accordance with all l...

Страница 10: ...anything Damage can occur to the table The PF400 Tables offer trendelenburg height backrest and footrest positioning with electronic ease Operate the foot control as shown below to raise or lower the...

Страница 11: ...or holding anything Damage can occur to the table The PF250 PF200 tables offer trendelenburg and height positioning with electronic ease Operate the foot control as shown below to raise or lower the t...

Страница 12: ...his area for holding anything Damage can occur to the table The PF400 Tables offer trendelenburg height backrest and footrest positioning with electronic ease Operate the optional hand control as show...

Страница 13: ...nder the backrest top is not designed for storage DO NOT use this area for holding anything Damage can occur to the table The PF250 PF200 tables offer trendelenburg and height positioning with electro...

Страница 14: ...use the table to transport a patient 11 DIRECTIONS FOR USE Grasp at base corner at low height range 1 2 13mm Open end Wrench Leveler Foot 1 Place the table in the location where you plan to use it 2...

Страница 15: ...side of the back rest by removing the screws and relocating the hand lever to the opposite side of the top into the four pre drilled holes Back Rest Control The Lowered Arm Rest can be removed when no...

Страница 16: ...CKING THE CASTERS 1 Place your shoe on top of the casters locking tab 2 Press firmly inward and downward until you here a click ing noise and the caster is unlocked CAUTION TABLE EXTENDER USE Do not p...

Страница 17: ...rm rest at a 60 angle to the backrest 2 Push end all the way into the hole and drop the arm rest into the horizontal position Pull outward to make sure the arm rest is secure TO LEVEL THE ARM RESTS Th...

Страница 18: ...ting for height and angle and then locking the cam locks into the desired position Important Safety Note Maximum Distance from edge of Table 3 7 6 cm QuickLock Face Rest Insert dowels into Accessory a...

Страница 19: ...cquer plated metals 2 Clorox Hydrogen Peroxide Disinfectant EPA reg 67619 24 corrosion of metals deterioration of upholstery and wood lacquer leaves a hard to remove residue 3 Virox Accel TB EPA reg 7...

Страница 20: ...ain wet for the disinfectant manufacturer s recommended contact time 4 Throw the used wipe s away and allow the table to dry INSPECTIONS RECOMMENDED REGULAR INSPECTIONS monthly or local standard Check...

Страница 21: ...Series Exam Table Exam Base Top FT Flat PF200 LB Lift Assist PF250 ES Electric PF400 Table Width Minimum Table Height Table Top Length Padding PL 3 AC 4 Fabric TerraTouch Grommets Options Model Number...

Страница 22: ...ICAL SPECIFICATIONS Designed for North America Europe Input Service 120 VAC 15 amp 60 Hz 220 VAC 10 amp 50 60 Hz Current Draw 4 0 amps 2 0 amps Maximum Momentary Current Consumption 6 0 amps 3 0 amps...

Страница 23: ...netic environment in which radiated RF disturbances are controlled the customer or the user of the table can help prevent electromagnetic interference by maintaining a minimum distance between portabl...

Страница 24: ...de 2 kV common mode 1 kV differential mode 2 kV common mode Mains power quality should be that of a typical com mercial or hospital environ ment Voltage dips short inter ruptions and voltage variation...

Страница 25: ...RF transmitters as deter mined by an electromagnetic site survey ashould be less than the compliance level in each frequency range b Interference may occur in the vicinity of equipment marked with th...

Страница 26: ...23 THIS PAGE IS INTENTIONALLY LEFT BLANK...

Страница 27: ...24 THIS PAGE IS INTENTIONALLY LEFT BLANK...

Страница 28: ...m PA 17349 Phone 717 235 6807 FAX 717 235 6798 www oakworksmed com made in the USA with US imported parts 3034177 CONFORMS TO ANSI AAMI STD ES60601 1 IEC STD 60601 1 3RD EDITION IEC STD 60601 1 2 3RD...

Отзывы: