SPECIFICATIONS
SPECIFICATIONS
29
Table 206 – Recommended Separation Distances between portable and mobile RF
Communications equipment and the CFPM400
Equipment and Systems that are NOT Life-supporting
Recommended Separations Distances for the CFPM400
The CFPM400 is intended for use in the electromagnetic environment in which radiated
disturbances are controlled. The customer or user of the CFPM400 can help prevent
electromagnetic interference by maintaining a minimum distance between portable and
mobile RF Communications Equipment and the CFPM400 as recommended below,
according to the maximum output power of the communications equipment.
Max Output Power
(Watts)
Separation (m)
150kHz to 80MHz
D=(3.5/V1)(Sqrt P)
Separation (m)
80 to 800MHz
D=(3.5/E1)(Sqrt P)
Separation (m)
800MHz to 2.5GHz
D=(7/E1)(Sqrt P)
0.01 .1166 .1166 .2333
0.1 .3689 .3689 .7378
1
1.1666 1.1666 2.3333
10
3.6893 3.6893 7.3786
100
11.6666 11.6666 23.3333
CFPM100 & CFPM400
table
table
table
SPECIFICATIONS