Longitudinal Travel
- Activates up to 10” of
horizontal top motion
away from base
Longitudinal Travel
- Activates up to 10” of
horizontal top motion
towards base
Right Side Tilt Control:
-
activates tilt up to 15º
Lateral Roll
Reverse Trendelenburg
-
activates tilt up to 12º
Reverse Trendelenburg
position
DIRECTIONS FOR USE
DIRECTIONS FOR USE
DIRECTIONS FOR USE
11
ADJUSTING TILT & LONGITUDINAL TRAVEL (CFPM400 ONLY)
The CFPM400 also comes with a hand control for controling
Trendlenburg, Lateral roll and Longitudinal travel.
Bubble levels are installed on the handle end of the table to aid the operator
to determine the horizontal plane.
Trendelenburg
-
activates tilt up to 15º
Trendelenburg
position
Lateral Roll
-
activates tilt up to 15º
Lateral Roll
2
3
4
1
Bubble Levels
PATIENT POSITIONING STRAP
The Patient Positioning strap must be used during all Trendelenburg/Reverse
Trendelenburg operations to prevent the patient from falling off the table. The
strap has Velcro
®
attached to the ends. Strap is attached as follows:
Patient Safety Strap must
be secured prior to using
the Trendelenburg
function.
CAUTION
1. Place the center of the strap around the
patient’s waist. Relocate the strap as
neccessary for adequate clearance.
2. Take the ends of the strap down around the
table top. Pull the strap snug underneath the
table and press the two Velcro
®
ends together.
3. Check the strap to assure adequate patient
restraint prior to the Trendelenburg motion.
5
6
There may be pinch
points under the table
top. Do not have any
body part under the
top when the table is
in motion.
CAUTION