Fluoroscopy Table

INSTRUCTION MANUAL

CFPM100

integrated headrest top

CFPM400

integrated headrest top

CFPMFXH

integrated headrest top

CFPMFXH

rectangular top

CFPM100

rectangular top

CFPM400

rectangular top

Содержание CFPM100

Страница 1: ...luoroscopy Table INSTRUCTION MANUAL CFPM100 integrated headrest top CFPM400 integrated headrest top CFPMFXH integrated headrest top CFPMFXH rectangular top CFPM100 rectangular top CFPM400 rectangular top ...

Страница 2: ...n contained in this document is subject to change without notice and should not be construed as a commitment by OAKWORKS Inc OAKWORKS Inc assumes no responsibility for any errors that may appear in this docu ment nor does it make expressed or implied warranty of any kind with regard to this material including but not limited to the implied warranties of merchantability and fitness for a particular...

Страница 3: ...00 400 only pg 10 Adjusting Tilt Longitudinal Travel CFPM400 only pg 11 Patient Positioning Strap pg 11 Adjusting Table Top Pad pg 12 Adjusting Cranial Positioning Pads integrated headrest only pg 12 Moving the Fluoroscopy Table pg 13 Emergency Stop pg 13 Usable Imaging Area Diagrams CFPMFXH CFPM100 integrated headrest tops pg 14 CFPMFXH CFPM100 rectangular tops pg 15 CFPM400 integrated headrest t...

Страница 4: ...air of stationary products All warranties exclude damage caused by improper set up accident abuse misuse neglect use for other than intended purpose or reasonable wear such as tears in the upholstery All warranties are invalidated by non factory modifications and unauthorized repairs which will immediately terminate all liability by OAKWORKS for the product or damages caused by its use The buyer a...

Страница 5: ...e B Applied Part pursuant to IEC 601 1 and EN 60601 1 1990 This symbol when used in this manual or on product labels indicates a Protective Earth Ground Terminal This symbol when used in this manual or on product labels warns that during trans port there should be no stacking of containers This symbol when used in this manual or on product labels indicates that the product should be protected from...

Страница 6: ...CFPMFXH is a fixed height table It is recommended that all patients are assisted onto the table physical assistance step stools with side rails is recommended Patient Safety Strap must be secured prior to using the Trendelenburg Reverse Trendelenburg function CFPM400 only INDICATIONS The OAKWORKS Fluoroscopy CFPM tables are indicated for use with mobile or compact C arm Fluoroscopy Imaging Systems...

Страница 7: ...h the x ray generator above the tabletop and a film cassette located on the tabletop The x ray generator should never be located above the tabletop when the OAKWORKS Fluoroscopy Table and the OAKWORKS Spine Positioning System Prone Pillow are used together This type of use requires that the x ray generator is located below the tabletop and the imaging intensifier or film cassette located above the...

Страница 8: ... TerraTouch Fabric 1 2 H x 12 SQ Head Rest Pad 1 4 H x 12 SQ Head Rest Pad 1 Crescent Face Pad Optional Accessories Carbon fiber arm board Spine Positioning System Fluoro Extender 2 Table top pad Carbon Fiber Top with Removable Table Top Pad Locking Casters 2 sets Integrated Face Rest with Face Cushion Hand Rail Includes 3 pad Set 1 2 H x 12 SQ Pad 2 4 H x 12 SQ Pad 3 Crescent Face Pad 1 2 3 The C...

Страница 9: ... 159 kg shipping 400 lbs 182 kg Included Accessories Table Top Pad with 1 Medical Grade Foam TerraTouch Fabric Optional Accessories Carbon fiber arm board Spine Positioning System Fluoro Extender 2 Table top pad Carbon Fiber Top with Removable Table Top Pad Locking Casters 2 sets Hand Rail The CFPMFXH is a fixed height table It is recommended that all patients are assisted onto the table physical ...

Страница 10: ...of Operation 120 VAC 15 amps 60Hz 5 8 amps 9 0 amps 120 VDC Class I Equipment Type B Applied Part Continuous Operation 10 Duty Cycle 220 VAC 10 amps 50Hz 3 0 amps 4 5 amps 220 VDC Class I Equipment Type B Applied Part Continuous Operation 10 Duty Cycle C U S L I S T E D C U S L I S T E D Table Specs Height Range 22 40 56 cm 102 cm Width 22 56 cm Length 84 214 cm Patient Capacity 500 lbs 227 kg dis...

Страница 11: ...amps 9 0 amps 120 VDC Class I Equipment Type B Applied Part Continuous Operation 10 Duty Cycle 220 VAC 10 amps 50Hz 3 0 amps 4 5 amps 220 VDC Class I Equipment Type B Applied Part Continuous Operation 10 Duty Cycle C U S L I S T E D C U S L I S T E D Table Specs Height Range 22 40 56 cm 102 cm Width 24 61 cm Length 80 203 cm Patient Capacity 500 lbs 227 kg distributed evenly Table Weight 350 lbs 1...

Страница 12: ...ss I Equipment Type B Applied Part Continuous Operation 10 Duty Cycle 220 VAC 10 amps 50Hz 4 5 amps 6 7 amps 24V Top 220V Column Class I Equipment Type B Applied Part Continuous Operation 10 Duty Cycle C U S L I S T E D C U S L I S T E D Table Specs Height Range 26 44 66 cm 112 cm Width 22 56 cm Length 84 214 cm Patient Capacity 500 lbs 227 kg distributed evenly Table Weight 475 lbs 216 kg shippin...

Страница 13: ...B Applied Part Continuous Operation 10 Duty Cycle 220 VAC 10 amps 50Hz 4 5 amps 6 7 amps 24V Top 220V Column Class I Equipment Type B Applied Part Continuous Operation 10 Duty Cycle C U S L I S T E D C U S L I S T E D Table Specs Height Range 26 44 66 cm 112 cm Width 24 61 cm Length 80 203 cm Patient Capacity 500 lbs 227 kg distributed evenly Table Weight 475 lbs 216 kg shipping 525 lbs 238 kg Con...

Страница 14: ...e use of this device create a circumstance under which the patient could be over exposed to the x ray being used discontinue use immediately and determine an alternative radiology table or alternative x ray generating source to use ADJUSTING HEIGHT OF THE TABLE CFPM100 400 The CFPM100 CFPM400 come complete with one foot control Operate the controls as shown below to raise or lower the height of th...

Страница 15: ...osition Lateral Roll activates tilt up to 15º Lateral Roll 2 3 4 1 Bubble Levels PATIENT POSITIONING STRAP The Patient Positioning strap must be used during all Trendelenburg Reverse Trendelenburg operations to prevent the patient from falling off the table The strap has Velcro attached to the ends Strap is attached as follows Patient Safety Strap must be secured prior to using the Trendelenburg f...

Страница 16: ...cent Face Pad and adjust accordingly Supine Positioning Use either the 2 or 4 square pad for proper placement ADJUSTING THE TABLE TOP PAD ADJUSTING THE CRANIAL POSITIONING PADS INTEGRATED HEADREST TABLES ONLY Underside of Tabletop Pads are held in place by hook loop fasteners The Crescent Face Pad can be adjusted as needed Place close in to the edge of the table with sides together for smaller pat...

Страница 17: ...er Table may be moved Push down until locking lever clicks into place Unlocked caster Locked Caster Table locked down Casters are designed for normal Hospital Clinic environments Not intended for outdoor travel use NOTE EMERGENCY STOP The CFPM100 CFPM400 are equipped with an emergency stop switch To stop the operation push the red button To reset rotate the red button clockwise ...

Страница 18: ...DIRECTIONS FOR USE USABLE IMAGING AREA DIRECTIONS FOR USE USABLE IMAGING AREA CFPMFXH CFPM100 INTEGRATED HEAD REST TOP TABLES ONLY 14 DIRECTIONS FOR USE ...

Страница 19: ...DIRECTIONS FOR USE DIRECTIONS FOR USE USABLE IMAGING AREA DIRECTIONS FOR USE USABLE IMAGING AREA CFPMFXH CFPM100 RECTANGULAR TOP TABLES ONLY 15 ...

Страница 20: ...DIRECTIONS FOR USE DIRECTIONS FOR USE USABLE IMAGING AREA 16 DIRECTIONS FOR USE USABLE IMAGING AREA CFPM400 INTEGRATED HEADREST TOP TABLE ONLY ...

Страница 21: ...DIRECTIONS FOR USE DIRECTIONS FOR USE USABLE IMAGING AREA CFPM400 RECTANGULAR TOP TABLE ONLY DIRECTIONS FOR USE USABLE IMAGING AREA 17 ...

Страница 22: ...t do not allow the patient to sit on the face rest area beyond the warning line on the table top WEIGHT CAPACITY PATIENT USE Casters should be locked when patient is mounting or dismounting the table Table should be flat when mounting and dismounting Patient should mount and dismount near the lifting column Table top should be moved fully to the rear after each use DO NOT sit beyond this point SAF...

Страница 23: ...GHT OR TILT CFPM100 CFPM400 MODELS Check the outlet to be sure that it has power and that the power cable is plugged in Check that the red emergency stop button is in the raised position If the table fails to operate contact our customer service department The Carbon Fiber table weighs as much as 475 lbs 216 kg Turning the table on its side is not required and not recommended CAUTION ...

Страница 24: ...et 6 Attempt to operate the table elevation functions from the Foot Control 7 If the table elevation functions fail to operate contact the Customer Service Department HAND CONTROL REPLACEMENT INSTRUCTIONS CFPM400 Replacement of the hand control will be necessary if the hand control does not activate the trendelenburg or lateral roll function To replace the hand control follow these steps 1 Before ...

Страница 25: ...orite solution phenol based surface disinfectants and quartenary ammonium compounds may be used Wipe off excess liquid and remove any residual solution with a damp not wet lint free cloth Be sure underside of table pad is completely dry prior to placing back onto tabletop The operator should ascertain the disinfecting properties of the agent being used prior to cleaning IMPORTANT IMPORTANT CAUTION...

Страница 26: ...e angle you need It will stay at that position until you change it When finished Ask the patient to raise up enough for you to move the Arm Board base from under the table top pad Fold it up and store it away until the next time you need to use it Carbon Fiber Arm Board base slides under table top pad Upper Arm Board rotates as needed a full 180º to 180º Prone Position at 135º Supine Position at 1...

Страница 27: ...r is radiolucent and offers extra width only where you need it for greater positioning accuracy with optimum imaging The Fluoro Extender is inserted under the table top pad CRESCENT FACE REST PAD T FACE CONTOURED TORSO SUPPORT PAD 7 X 12 RECTANGULAR ADJUSTER PADS 2 8 X 16 RECTANGULAR ADJUSTER PADS 2 CONTOURED TORSO WEDGES 2 RADIOLUCENT FRAME WITH ADJUSTABLE FACE REST CARRY CASE 8 HALF ROUND BOLSTE...

Страница 28: ...2280CF CFPM100 Carbon Fiber Rectangular Top Continental Plug Europe FLDBTLETEU222684CF CFPM400 Carbon Fiber Integrated Headrest Top Continental Plug Europe FLDBTLRTETEU242680CF CFPM400 Carbon Fiber Rectangular Top Continental Plug Europe FLDBFD223784CF CFPMFXH Carbon Fiber Integrated Headrest Top FLDBRTFD243780CF CFPMFXH Carbon Fiber Rectangular Top FLDBSW222284CF CFPM100 Carbon Fiber Integrated H...

Страница 29: ...5 POWER CORD VARIABLES Plug View Voltage AC Power Cord Plug 120v 60 Hz 3 prong grounded hospital grade North America 220v 50 Hz Continental Plug Europe 220v 50 Hz Swiss plug Switzerland 220v 50 Hz British plug United Kingdom Power Cord Variables ...

Страница 30: ...ted Face Rest Pad 4 H x 12 SQ 1 2410 06 Crescent Face Pad 9468 06 1 Inch Tabletop Pad 22 x 84 Integrated Head Rest Top 58940 06 1 Inch Tabletop Pad 24 x 80 Rectangular Top Optional Accessories 3395 01 Carbon Fiber Arm Board 1 4796 06 2 5cm x 12 31cm Square Support Pad 1 3PXXXXXXXXTT OAKWORKS Prone Pillow 2005 Fluoro Extender 8806 Patient Positioning Strap 6427 06 2 Inch Table Top Pad 22 x 84 Integ...

Страница 31: ...letop AL mm Materials of Construction Tabletop Patient Comfort Carbon Fiber 1 64 x 22 x 84 41 7 mm x 559 mm x 2134 mm 1 25 mm upholstered foam pad This product contains no latex Complies with UL 60601 1 CSA C22 2 601 1 IEC 60601 1 Ed2 IEC 60601 2 32 Ed1 IEC 60601 1 2 Ed2 1 CFPM100 400 only electromagnetic compatibility 03 08 ETL cETL Listed Storage Transport Temperature 32 100º F 0 38ºC Humidity 6...

Страница 32: ...nce Electromagnetic Environment Guidance RF Emissions CISPR 11 Class B Group 1 The CFPM400 uses RF energy only for its internal function Therefore its RF emissions are very low and are not likely to cause any interference in nearby electronic equipment Harmonics IEC 61000 3 2 Class A Flicker IEC 61000 3 3 Complies The CFPM400 is suitable for use in all establishments including domestic and those d...

Страница 33: ...user of the CFPM400 can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF Communications Equipment and the CFPM400 as recommended below according to the maximum output power of the communications equipment Max Output Power Watts Separation m 150kHz to 80MHz D 3 5 V1 Sqrt P Separation m 80 to 800MHz D 3 5 E1 Sqrt P Separation m 800MHz to 2 5G...

Страница 34: ...pical commercial or hospital environment Surge EN IEC 61000 4 5 1kV Differential 2kV Common 1kV Differential 2kV Common Mains power quality should be that of a typical commercial or hospital environment Voltage Dips Dropout EN IEC 61000 4 11 95 Dip for 0 5 Cycle 60 Dip for 5 Cycles 30 Dip for 25 Cycles 95 Dip for 5 Seconds 95 Dip for 0 5 Cycle 60 Dip for 5 Cycles 30 Dip for 25 Cycles 95 Dip for 5 ...

Страница 35: ...IEC 61000 4 6 Radiated RF EN IEC 61000 4 3 3 Vrms 150 kHz to 80 MHz 3 V m 80 MHz to 2 5 GHz 3 Vrms 3 V m Portable and mobile communications equipment should be separated from the CFPM400 by no less than the distances calculated listed below D 3 5 V1 Sqrt P D 3 5 E1 Sqrt P 80 to 800 MHz D 7 E1 Sqrt P 800 MHz to 2 5 GHz where P is the max power in watts and D is the recommended separation distance i...

Страница 36: ... S A CONTACT INFORMATION OAKWORKS Medical Equipment 923 East Wellspring Road New Freedom PA 17349 Toll Free USA only 800 916 4612 Phone 717 235 6807 Toll Free FAX USA only 877 562 4787 FAX 717 235 6798 www oakworksmed com European Compliance for CE Mark Information Emergo Europe Molenstraat 15 The Hague 2513 BH Netherlands Tel 31 70 3458570 Fax 31 70 3467299 E mail info emergogroup com ...