- 19 -
There are two calibration options:
Offset – adjusts at a single point. Offset
calibration can be performed at any
temperature in the offset range of ±10.00
°C or 18.00 °F
Slope – adjusts at two points. The two
calibration points must be at least 40.00 °C
(72.00 °F) apart. The 2nd calibration point
should be at a higher value than the 1st
calibration point.
Use the
▲
or
▼
keys to adjust the value to
match known temperature standard. Press
F1
to accept.
www.calcert.com
1.888.610.7664
0
5
10
15
20
25
30