- 10 - 

8. KEY 

FUNCTIONS 

 

 

 

Note: Function keys change in setup 

mode to provide advanced operation 

flexibility. Text above key will indicate 

function. 

F1

 

Step through Min, Max and 
Avg readings. 

F2

 

Toggle between F and C 
display 

F3

 

Toggle between menu and 
measure mode 

hold

 

Freeze display 

on/off 

light

 

Turns meter on and off (press 
and hold for 3 seconds to turn 
off) 
Press momentarily to turn on 
backlight  

recall

 

Recalls and steps through 
stored readings 

log

 

Stores current measured 
value to memory 

www.calcert.com

[email protected]

1.888.610.7664

0

5

10

15

20

25

30

Содержание WD-35426-60

Страница 1: ...Instruction Manual Temp 360 RTD Datalogging Thermometer 68X526601 Rev 1 09 09 Part of Thermo Fisher Scientific www calcert com sales calcert com 1 888 610 7664 0 5 10 15 20 25 30...

Страница 2: ...MODE 13 11 HOLD FUNCTON 14 12 MIN MAX and AVE FUNCTION 14 13 DATA LOGGING 14 14 SETUP MODE 15 15 GENERAL SETUP SCREEN 16 16 CALIBRATION SCREEN 18 17 ALARMS SCREEN 20 18 DATA LOGGING SCREEN 25 19 CALI...

Страница 3: ...SB output Operator selection of Celsius or Fahrenheit scale Resolution of up to 0 01 C F Large backlit LCD and dot matrix graphic LCD display Hold feature for temporarily retain a reading Displays min...

Страница 4: ...only be changed in an area known to be non hazardous CAUTION 1 Do not use or store this instrument in microwave ovens or any abnormally hot or cold areas 2 Weak batteries should not be left in the ins...

Страница 5: ...C F From 100 0 to 999 9 C F 1 C F Above 1000 C F Accuracy 0 1 C 0 2 F From 330 0 to 100 C F 0 03 C 0 06 F From 99 99 to 99 99 C F 0 1 C 0 2 F From 100 0 to 999 9 C F 1 C 2 F Above 1000 C F Display Ba...

Страница 6: ...50V rms Storage 40 C to 65 C 40 F to 149 F Humidity 10 to 90 non condensing Battery Life Size Three AA 1 5V Alkaline Life 400 hours continuous typical without backlighting and buzzer on Dimensions Wi...

Страница 7: ...ess protection Meets IEC 529 IP 54 for dust and water resistant enclosures probe attached CE Compliance EN61326 1 A1 1998 EU EMC Directive www calcert com sales calcert com 1 888 610 7664 0 5 10 15 20...

Страница 8: ...V Adaptor power supply Main power supply connected USB power supply USB power Selected settings are stored in memory and will remain in memory even after power is turned off or while batteries are bei...

Страница 9: ...eter before pushing the bottom edges of meter down to set it into position Lift up the stand at the back of meter for bench top applications if necessary 2 To remove thermometer from armor push out fr...

Страница 10: ...EMBLING OPTIONAL HANDSFREE ACCESORIES You can use the optional magnets and strap in the Handsfree Kit accessories for hands free operations www calcert com sales calcert com 1 888 610 7664 0 5 10 15 2...

Страница 11: ...e mating connector on the top of the instrument If no probe is connected the display will read open RTDs are sensitive at the tip or sensing element When taking measurements allow time for the reading...

Страница 12: ...ings F2 Toggle between F and C display F3 Toggle between menu and measure mode hold Freeze display on off light Turns meter on and off press and hold for 3 seconds to turn off Press momentarily to tur...

Страница 13: ...ot matrix display features a large primary display smaller secondary displays for channel info or min max ave and helpful annuciators for added measurement data www calcert com sales calcert com 1 888...

Страница 14: ...PC data logging indicator 10 Main reading display 11 Current reading unit indicator 12 Min Max Avg elapsed time 13 Current Min Max Avg reading indicator 14 Meter logging memory location 15 Measurement...

Страница 15: ...or in the primary display Pressing the F2 key will toggle reading between F and C display Pressing the F1 key initiates and toggles through Minimum Maximum and Average reading modes Pressing F3 enter...

Страница 16: ...and maximum values are sensed and automatically stored To exit and clear this function press the F3 to access the Menu functions See the Clear Reset menu section for more details 13 DATA LOGGING Press...

Страница 17: ...options To enter a setup screen press Select F1 key To return to measurement mode press Meas F3 key Following menu options are listed 1 General Setup 2 User field calibration 3 Alarm settings 4 Data l...

Страница 18: ...ss F1 to indicate you want to change the setting of the current parameter or recall or log to move to the next parameter Press recall or log to change the options Press F2 to choose the next setting W...

Страница 19: ...ed to reset change password In the event if uses forget his her password 5586 can be used to reset to a new value On the second page you can set time and date www calcert com sales calcert com 1 888 6...

Страница 20: ...compensate for RTD off set error It is NOT necessary to perform a field calibration to obtain specified meter accuracy Use the field calibration feature to improve thermometer probe accuracy or to com...

Страница 21: ...0 C or 18 00 F Slope adjusts at two points The two calibration points must be at least 40 00 C 72 00 F apart The 2nd calibration point should be at a higher value than the 1st calibration point Use th...

Страница 22: ...nder the alarm setting options Measurement alarm Disable or enable the alarm by pressing recall or log and F1 to accept Increase or decrease individual limit by pressing recall or log Meter in alarm m...

Страница 23: ...e disable the countdown alarm and set the countdown time from 5 sec to 1 hour After setting enabling the countdown alarm the measurement screen should look like this www calcert com sales calcert com...

Страница 24: ...e of a countdown will stop the process You will need to restart the countdown by selecting the start key F2 Alternatively you can also choose Menu to go into the timer menu and disabling the alarm www...

Страница 25: ...u display should look like this To repeat the functions select F2 Note The C F function is disabled when the count down function is enabled To activate the o C o F function on measurement screen pleas...

Страница 26: ...ff until the countdown is completed or manually stopped If the meter is manually or auto shutoff The Countdown Alarm is automatically reset to Disable each time the meter is shut off The Timer key wil...

Страница 27: ...log to set time interval Its range is from 2min to 60min Data Transfer from Meter to Computer Once the USB connection is establish with PC press the Select button to download data from Meter to PC us...

Страница 28: ...PORT SCREEN The Calibration report will show the time and date along with results of the last user calibration www com information itm com 1 800 561 8187 www calcert com sales calcert com 1 888 610 76...

Страница 29: ...a you want to clear or reset For calibration logged data and reset all you will have to enter the password to proceed Default Password is 9900 www com information itm com 1 800 561 8187 www calcert co...

Страница 30: ...clean with abrasives or solvents Use mild detergents never immerse nor use excessive fluid 22 BATTERIES If there is no display when the thermometer is turned on check condition of the three AA batter...

Страница 31: ...eck that batteries are inserted properly Display shows OVER UNDER Out of range indication Display Shows OPEN No RTD connected in the Connector Display Shows Err If display shows this message other tha...

Страница 32: ...ndsfree Kit Two Magnets and a Strap 35427 85 General purpose probe immersion Into liquids 08117 70 Penetration probe meat semi soft Materials 08117 85 Surface probe direct contact on Hot surfaces 0811...

Страница 33: ...oduct misuse or abuse may be invoiced to you 26 PRODUCT RETURN To limit charges and delays contact the seller or Manufacturer for authorization and shipping instructions before returning the product e...

Страница 34: ...ndards Conformity ISO IEC 17025 2005 accredited NIST Handbook 150 2000 Edition ANSI NCSL Z540 2 1997 NIST Technical Note 1297 ISO 9000 2000 Fast Service Our substantial inventory of replacement parts...

Страница 35: ...n you need to meet your most stringent quality requirements for the control of inspection measuring and test equipment Certification includes certificate of calibration with test data including descri...

Страница 36: ...ibrate your instrument Are your calibration certificates good enough InnoCal surpasses the competition by providing the most complete certificates as required by NIST All of our certificates include m...

Страница 37: ...traceable standards Meter only Probe only System meter probe Four test points across range of instrument 0 100 165 230 C 4 32 446 770F MM 17000 04 MM 17001 04 MM 17002 04 www calcert com sales calcer...

Отзывы: