background image

- 1 - 

1. INTRODUCTION 

 

This versatile hand-held 

instrument provides highly 

accurate temperature 

measurements. The 

instrument is designed    

for easy operation and 

includes

 

the following 

features:

 

 

 

  Operator   selection   of   Celsius   or 

Fahrenheit scale 

  Resolution   of   0.1° C/F  from   -199.9 to 

999.9° 

  Large backlit LCD with two lines of 

four-digit display 

  Hold feature for temporarily retaining a 

reading 

  Displays min and max readings 

  Field calibration capability 

  Disabling of Auto-Off function  

  Low battery warning 

  Two blade female ANSI mini- 

connector input 

  Operates with a wide selection of 

probes 

Содержание Temp10J

Страница 1: ...INSTRUCTION MANUAL Thermocouple Thermometer Type J K T Part of Thermo Fisher Scientific 68X450501 Rev 0 02 08 ...

Страница 2: ......

Страница 3: ...THERMOCOUPLE 8 9 KEY FUNCTIONS 10 10 Display Overview 11 11 SELECTING TEMPERATURE SCALE12 12 HOLD FUNCTON 12 13 MIN AND MAX FUNCTION 13 14 AUTO OFF FUNCTION 13 15 CALIBRATION 13 16 CANCELING A FIELD CALIBRATION14 17 FIELD CALIBRATION LOCKOUT AND RE ENABLE 15 18 MAINTENANCE 16 19 CLEANING 16 20 BATTERIES 16 21 TROUBLE SHOOTING 17 22 ACCESSORIES 18 23 WARRANTY 19 24 PRODUCT RETURN 19 25 INNOCAL CALI...

Страница 4: ...perator selection of Celsius or Fahrenheit scale Resolution of 0 1 C F from 199 9 to 999 9 Large backlit LCD with two lines of four digit display Hold feature for temporarily retaining a reading Displays min and max readings Field calibration capability Disabling of Auto Off function Low battery warning Two blade female ANSI mini connector input Operates with a wide selection of probes ...

Страница 5: ...ous atmosphere batteries must only be changed in an area known to be non hazardous CAUTION 1 Do not use or store this instrument in microwave ovens or any abnormally hot or cold areas 2 Weak batteries should not be left in the instrument Dead batteries can leak and cause damage to unit DANGER 1 Voltages present at the thermocouple may also be present at the battery terminals Always disconnect the ...

Страница 6: ...T 418 F to 752 F Out of rangedisplay Resolution 0 1 1 auto ranging 0 1 C F from 199 9 to 999 9 1 C F outside this range Accuracy Below 150 C 238 F 0 25 of reading 1 C 2 F Above 150 C 238 F 0 1 of reading 0 4 C 0 7 F Display Backlit LCD of viewing area 58 X 40mm Min Max Function Yes Auto Off after 17 5 minutes Enable Disable option available Stability Criteria Yes upon stability of 5 seconds ...

Страница 7: ...ttery Life Size Three AA 1 5V Alkaline Life 750 hours continuous typical without backlighting Dimensions Without Armor 175mm L X 97mm W X 42mm H With Armor 180mm L X 102mm W X 52mm H Weight with batteries Without Armor 267g With Armor 362g Ingress protection Meets IEC 529 IP 54 for dust and water resistant enclosures probe attached CE Compliance EN61326 1 A1 1998 EU EMC Directive ...

Страница 8: ...icates 150 hours remaining Selected settings are stored in memory and will remain in memory even after power is turned off or while batteries are being replaced 1 Before changing battery turn instrument off and disconnect thermocouple 2 Loosen screw and lift battery cover off back of case 3 Remove the three AA batteries 4 Insert three new batteries observing polarity 5 Install cover and tighten sc...

Страница 9: ...er armor slide in from the top of meter before pushing the bottom edges of meter down to set it into position Lift up the stand at the back of meter for bench top applications if necessary 2 To remove thermometer from armor push out from the bottom edges of meter until it is completely out of boot ...

Страница 10: ... 7 6 ASSEMBLING OPTIONAL HANDSFREE ACCESORIES You can use the optional magnets and strap in the Hands Free Kit accessories for hands free operations ...

Страница 11: ...ECTING A THERMOCOUPLE Use the correct thermocouple type for your instrument Using an incorrect thermocouple type will result in erroneous readings Thermocouples are color coded by type using the North American ANSI Color Code as follows TYPE COLOR J Black K Yellow T Blue Thermocouple connectors have one wide blade and one narrow blade Do not force connector in backwards Connect thermocouple to rec...

Страница 12: ...rmocouple wiring polarity must be correct If readings decrease as the temperature increases the thermocouple wires may be reversed The red wire is negative for thermocouple wires manufactured in North America ...

Страница 13: ...tivates Deactivates freezing of the measured reading while in measurement mode min max Activates Deactivates display of the minimum and maximum readings on off light Powers on and shuts off the thermometer by holding it for 3 seconds Otherwise it activates deactivates the backlight display Backlight will automatically turn off within 30 seconds of activation ...

Страница 14: ...urs 2 Bar 550 hours 1 Bar 350 hours Blinking 150 hours CAL Blinks during calibration mode Remains in display indicating field calibration is active HOLD Remain in display during hold mode STABLE Displayed upon recognizing final value Err CAL Error indicator on main digit display Indicates calibration error due to out of tolerance value ...

Страница 15: ...ll give you the approximate time required 11 SELECTING TEMPERATURE SCALE Select C or F by pressing the C F key Each time the key is pressed the temperature scale will switch Switching between C and F can be done at any time during operation Each time you turn the instrument on it will power up with the same settings that were set when the unit was last turned off 12 HOLD FUNCTON Press the hold key...

Страница 16: ...ter has a default auto off function of 17 5 minutes To deactivate this function press and hold the min max key while turning on the meter using the on off key A OFF nO will appear indicating that the auto off function is disabled 15 CALIBRATION The thermometer is factory calibrated and does not require calibration before use The Calibration function allows single point calibration of the thermomet...

Страница 17: ... 50 F when the temperature reading is stable press the cal key The CAL indicator will stop blinking and the reading will be set to 0 C 32 F The CAL indicator will remain turned on indicating a field calibration is active If Err CAL is displayed either the displayed reading is outside the above limits or the batteries are weak 16 CANCELING A FIELD CALIBRATION 1 Turn the thermometer off 2 Hold the c...

Страница 18: ...ation operation LOCKOUT PROCEDURE 1 Turn the thermometer off 2 Simultaneously press and hold the cal and the C F keys down and momentarily press the on off key Continue to hold the cal and C F keys until the measurement mode is displayed RE ENABLE PROCEDURE 1 Turn the thermometer off 2 Simultaneously press and hold the hold and the cal keys down and momentarily press the on off key Continue to pre...

Страница 19: ... BY ELECTROSTATIC DISCHARGE CLEAN WITH DAMP CLOTH Do not clean with abrasives or solvents Use mild detergents never immerse nor use excessive fluid 20 BATTERIES If there is no display when the thermometer is turned on check condition of the three AA batteries Also check that the battery terminals are clean and batteries are properly installed If replacement is necessary refer to the BATTERY INSTAL...

Страница 20: ...ts FAULT ACTION NO Display Check condition of batteries Check that batteries are inserted properly Display shows Out of range indication Display Shows OPEN No thermocouple connected in the Connector Display Shows Err If display shows this message other than during the field calibration mode please return the instrument for servicing ...

Страница 21: ...ds type J 08517 55 EC TPGLPJ 01M Penetration probe meat semi soft Materials type J 08517 65 EC TPPENJ 01M Surface probe direct contact on Hot surfaces type J 08517 60 EC TPSURJ 01M Clip on probe surface contacts Electronics type J 08469 00 EC TPCLPJ 01M General purpose probe immersion Into liquids type K 08516 55 EC TPGLPK 01M Penetration probe meat semi soft Materials type K 08516 65 EC TPPENK 01...

Страница 22: ...ts outside the warranty period or those resulting from product misuse or abuse may be invoiced to you 24 PRODUCT RETURN To limit charges and delays contact the seller or Manufacturer for authorization and shipping instructions before returning the product either within or outside of the warranty period When returning the product please state the reason for the return For your protection pack the c...

Страница 23: ...s of customers meet ISO FDA EPA GLPs cGMPs and other quality standards Conformity ISO IEC 17025 2005 accredited NIST Handbook 150 2000 Edition ANSI NCSL Z540 2 1997 NIST Technical Note 1297 ISO 9000 2000 Fast Service Our substantial inventory of replacement parts ensures a fast turnaround and prevents costly downtime Most instruments serviced in five business days Excellent Value Get quality at a ...

Страница 24: ...ice your instrument We provide you with the documentation you need to meet your most stringent quality requirements for the control of inspection measuring and test equipment Certification includes certificate of calibration with test data including description and identification of the item certified condition of the item issue date identification of calibration procedure calibration date as foun...

Страница 25: ...l send you an automatic reminder when it is time to recalibrate your instrument Are your calibration certificates good enough InnoCal surpasses the competition by providing the most complete certificates as required by NIST All of our certificates include measured data and point by point measurement uncertainty and by request we ll provide test accuracy and test uncertainty ratios at no extra cost...

Страница 26: ...Meters 270 to 2316 C 454 to 4200 F Probes Systems 80 to 1000 C 112 to 1832 F Actual range is dependent on type of probe MM 17000 10 MM 17001 10 MM 17002 10 InnoCal The Choice of Quality 866 InnoCal 866 466 6225 InnoCalSolutions com For calibration services outside of North America please contact your Local Distributors or Local Certification Body ...

Страница 27: ......

Страница 28: ...utor or visit our web site listed below OAKTON Instruments P O Box 5136 Vernon Hills IL 60061 USA Tel in U S 888 462 5866 Tel outside U S 1 847 549 7600 Fax 1 847 247 2984 Website www 4oakton com E mail info oakton com Thermo Scientific Water Analysis Environmental Instruments Blk 55 Ayer Rajah Crescent 04 16 Singapore 139949 Tel 65 6778 6876 Fax 65 6773 0836 Website www thermo com Email eutech th...

Отзывы: