4 MEASUREMENT
This meter is capable of taking measurements with automatic or manual temperature compensation.
Automatic temperature compensation only occurs when a temperature sensor is plugged into the
meter. If there is no temperature sensor plugged into the meter, the default manual temperature
setting is automatically 25
°
C. You can manually set the temperature to match your working
conditions using a separate thermometer.
4.1 Automatic
Temperature
Compensation
For automatic temperature compensation (ATC) simply plug
the temperature probe into the meter (see page 6 for
directions). The ATC indicator will light up on the LCD. See
figure 13.
NOTE: If you are using a temperature probe, the
probe must be submersed in the liquid you
are measuring.
Figure 13 - ATC annunciator will light
up when connected to temperature
probe