Instruction Manual
CyberScan pH 11 / 110
39
9.4
P3.0: View Electrode Data
Program 3 has two “view only” options that let you check the electrode parameters
for diagnostic purposes.
•
From pH measurement mode, you can view the electrode’s offset and slope
values.
•
From mV measurement mode, you can view the electrode’s relative mV offset
value. (pH 110 only)
9.4.1 View pH Electrode Offset and Slope
From pH measurement mode:
1. Press the SETUP key to enter Set Up mode.
2. Press
the
MI/
▲
or MR/
▼
keys to scroll through
subgroups until you view parameter P3.0.
3. Press the HOLD/ENTER key.
4. The display shows the electrode offset value. It
is the mV offset at pH 7.00. If you have not
calibrated at any buffer, the primary display
shows ‘---‘.
5. Press the HOLD/ENTER key to proceed to
electrode slope display.
6. The display shows electrode slope in
percentage. Slope displayed will be the current
slope of the slope zone to which a measurement
is made or calibration is done.
7. At any point, you can press the CAL/MEAS key
to return to measurement mode.
See Figure 35.
Figure 35: P3.1/3.2 - View pH
electrode offset and slope
mV
pH
SETUP
5.3
p3.1
SETUP
p 3.0
ele
%
pH
SETUP
p 3.2
100.0
ENTER
HOLD
ENTER
HOLD