Instruction Manual
CyberScan pH 11 / 110
22
7.2 Memory
Recall
This function recalls the previous readings stored in the
memory. You can only access MR from the
measurement mode. Memory recall is in “Last In First
Out” order. To recall readings:
1. Press the MR/
▼
key once to retrieve the last
reading stored. The memory location screen –
MEM, “Loc” and the memory number – will flash
on the display.
2. Press the HOLD/ENTER key to recall the
reading stored under that memory number.
3. To display next stored reading, press
HOLD/ENTER key. The display automatically
moves to the next lower memory location
screen.
4. To view a stored reading in a specific memory
location, press MI/
▲
or MR/
▼
keys to locate it,
and press HOLD/ENTER key to display the
reading.
NOTE: Readings stored in memory are retained even if
the unit is turned off. To erase all readings stored in
memory, use the SETUP in section 9.7 – P 6.0: Memory
Clear (CLr) on page 44.
MEM
MEM
loc
loc
25
26
MI
MR
Figure 21: Memory recall
Figure 22: Scroll up or down
for specific memory location