background image

Batteries are weak and need replacement soon.

stable error

Appears when calibration is attempted but the reading  

is not yet stable. Wait for the reading to stabilize or 

manually confirm the calibration by pressing enter.

buffer error The buffer is outside of the calibration range.
slope error The 2nd and 3rd calibration point is not within 80% to  

120% slope range.

over range The reading is above the measuring range of tester.

under range The reading is below the measuring range of tester.

Self-Diagnostic Messages

Replacing the Batteries

The PCTSTestr 50 Pocket Tester uses four AAA 1.5 V batteries.

1.  To remove the battery

cover, see Figure 1.

Clear the front catch

and then the back

catch, before sliding

the cover off.

2.  To remove the battery

plate, push the center

tab towards the front

of the tester as shown

in Figure 2. Once

unlocked, remove the

plate to access the

batteries.

3.  Invert the tester

upside down to

remove the batteries.

Each side uses two

AAA batteries. Orient

each battery with

positive terminal facing

downward.

4.  To lock the battery

place, align the small

tabs (Figure 3) into the

guide ribs on the hous-

ing and then press

down. See Figure 4.

Fig. 3: Align tabs

Fig. 4: Push down to lock

Fig. 5: Battery plate unlocked

Fig. 6: Battery plate locked

Fig. 2: Push to unlock

Fig. 1: Removing battery cover

User Reset

Factory  

Reset

Yes

No

Measure
Settings
Temp Set

hold

menu

hold

Reset

hold

1.  Press HOLD/

   

button to

freeze  the  measurement.

2.  Press HOLD/

   

again to

release the measurement.

hold

hold

COND

µs/cm

820 

23.5°C

Press Hold  

to release

COND

µs/cm

820

 

23.5°C

7

8

9

1065O100_MAN_35634-35 

August, 2017  

Ordering 

Code

Product Description

35634-35

PCTSTestr 50 pocket tester with case, lanyard,  

and batteries

35634-37

Replacement sensor module for PCTSTestr 50 

35634-09

Replacement sensor cap

09376-00

Replacement alkaline batteries; AAA, 1.5 V.  

Pack of 12

17101-45

NIST-traceable calibration with data for pocket testers

Accessories

Specifications

PCTSTestr 50 

pH

pH range

–1.00 to 15.00 pH

Resolution

0.01 pH

Relative accuracy

±0.01 pH

Calibration points

Up to 3 points

Buffer set standard 

selection

USA: 4.01/7.00/10.01

NIST: 4.01/6.86/9.18

Calibration window

±1.00 pH

Calibration type

Point to point

Conductivity

Conductivity range

0.0 to 200.0 μS, 200 to 2000 μS,  

2.00 to 20.00 mS

Resolution

0.1 μS, 1 μS, 0.01 mS

Relative accuracy

±1% full scale

Normalization temperature  25.0ºC (77ºF)

Temperature co-efficient

0.0% to 10.0%

Calibration points

Up to 3 points

TDS

TDS range

0.0 to 100.0 ppm, 100 to 1000 ppm,  

0.10 to 10.00 ppt (TDS factor 0.5)

Resolution

0.1 ppm, 1 ppm, 0.01 ppt

Relative accuracy

±1% full scale

Calibration points

Up to 3 points

TDS factor

0.40 to 1.00 (selectable)

Salinity

Salinity range

0.00 to 10.00 ppt  

Resolution

0.10 ppt

Relative accuracy

±1% full scale

Calibration points

1

www.4oakton.com

Specifications (cont.)

PCTSTestr 50 

Temperature

Temperature range

0 to 60°C (32.0 to 140.0°F)

Temperature resolution

0.1°C / 0.1°F

Temperature accuracy

From 0 to 50ºC (±0.5°C / ±0.9°F + 1 LSD);  

from 50 to 60°C (±1.0ºC / ±1.8°F + 1 LSD)

Temperature 

compensation

Yes (Automatic Temperature 

Compensation)

General

Display

Graphics, dot matrix 80 x 100 pixel

Backlight

Yes, selectable  

(30 sec from last key press)

Auto off

8.5 minutes (from last key press)

Reset

User / Factory

Power requirement

Four AAA 1.5 V batteries

Battery life

>150 hours

Water proofing

IP67

Regulatory  

certifications

CE, FCC

Environmental Operating Conditions

Ambient operating  

temperature

5 to 45°C / 41 to 113°F 

Relative humidity

5% to 85% noncondensing

Storage temperature

–20 to 60°C / –4 to 140°F

Storage humidity

5% to 85% noncondensing

10

11

Factory Reset

Reset to the Factory Default Settings by using the Factory 

Reset function. 
1.  Press MENU/

v

 

to enter setup window. Scroll down by

pressing MENU/

v

 

to select Reset. Press HOLD/

   

to select

Reset. The display shows User Reset and Factory Reset.

2.  Scroll down by pressing  MENU/

v

 

to toggle between the

resets. Press HOLD/

   

to select Factory Reset.

3.  The display automatically shows No and Yes. Scroll down

by pressing MENU/

v

 

to toggle between No and Yes.

4.  Press HOLD/

   

to confirm either No or Yes. The display

shows the Factory Reset option with a  .

HOLD Function

This feature lets you freeze the display for a delayed 

observation.

Sensor Maintenance

1.  Always keep the sensor electrodes clean. Rinse the

electrodes with de-ionized water and wipe them dry with

clean cloth before storing with its protective cap. For cup

type electrodes, remove the white plastic cup and insert

to thoroughly clean viscous solutions. Note: Never scratch

electrodes with a hard substance.

2.  For better performance, soak the electrode in alcohol for

10 to 15 minutes and rinse with de-ionized water before

starting any measurement process. This is to remove dirt

and oil stains on the electrode, which may affect the

accuracy of the measurements.

Sensor Replacement

You can replace the sensor module at a fraction of the cost 

of a new tester. When the tester fails to calibrate or gives 

fluctuating readings in calibration standards, you need to 

change the electrode.

1.  With dry hands, grip the ring with sensor facing you. 

Twist the ring counterclockwise. Save the ring for later use.

2. Pull the old sensor module away from the tester.
3.  Align the four tabs on the new module so that they

match the four slots on the tester.

4.  Gently push the module onto the slots to sit it in position.

Push the smaller O-ring fully onto the new sensor

module. Push the other O-ring over the module and

thread it into place by firmly twisting clockwise.

Note: 

It is necessary that you recalibrate your tester prior 

to measurement after a sensor replacement.

Warranty

This instrument is supplied with a warranty against  

manufacturing defects for a period of one year from the 

date of purchase.

Return of Items

Authorization must be obtained from your distributor  

before returning items for any reason. When applying for 

authorization, please include information regarding the  

reason the item(s) are to be returned.

We reserve the right to make improvements in design, 

 

construction and appearance of products without notice. 

Prices are subject to change without notice.

Specifications

Key Info   

Scroll 

Accept 

EXIT   

MENU 

HOLD 

CAL  

Press Hold  

to release

COND

µs/cm

820

 

23.5°C

COND

µs/cm

820 

23.5°C