5.  Measurement

Measuring Ion Concentration

1. 

Switch on meter. The ION 5 and ION 6 meters
will automatically enter the ion measurement
mode. No annunciator will appear.

See figure

2. 

Prepare samples as necessary (i.e., add Ionic
Strength adjuster). Sample preparation varies
depending on ion type—see your electrode manual for details on the specific
electrode that you are using.

3. 

Dip electrode into the sample. The sensor or glass bulb of the electrode must be
completely immersed into the sample. For best results, use a magnetic stirrer to
ensure a homogenous sample.

4. 

Wait for reading to stabilize. Note reading.

ION TIP: 

Concentration readings are not temperature compensated. Try to avoid

temperature fluctuations in your standards and samples.

Measuring mV

The mV mode is provided to allow you to check performance and calibration of
your ion selective electrode. 

To verify electrode operation, check mV reading in calibration standards one decade
apart in value (i.e.: 1 ppm; 10 ppm). Use the MODE key to toggle between ion and
mV measurement (ION 6 meter MODE key switches between pH, temperature, ion
and mV).

Consult your electrode manual for expected mV (slope) value. If your actual mV
readings are not within expected values as stated in your electrode manual, you
may need to recondition or replace your electrode.

5.1

5.2

A

A

Measuring pH and temperature (ION 6 meter only)

The Acorn ION 6 meter also takes readings in 
pH and temperature. 

1. 

Rinse both electrode and temperature sensor
with distilled water.

2. 

Turn meter on. ION 6 meter automatically
enters ion measurement mode. Press MODE
key  until meter enters pH measurement mode. 
pH annunciator will appear in the upper right
corner of the LCD.

See figure

3. 

Dip electrode and temperature sensor 
approximately 1" to 2" into sample. 
Stir gently and wait for display to stabilize.
Note pH reading.

4. 

Press MODE key to see your temperature 
reading. °C annunciator will appear in the lower right corner of the LCD.

See figure

Measuring pH with manual temperature compensation

When no temperature probe is connected, the meter will compensate based on a
manually entered temperature value. The factory default setting is 25°C (no temper-
ature compensation). You can compensate for temperature manually by entering the
temperature of your solution using the procedure below.

1. 

Press the MODE key until ”°C” shows in the display to enter Temperature mode.

2. 

Check the temperature of your solution using an NIST-certified thermometer or
other thermometer known to be accurate

3. 

Press CAL key to enter temperature calibration mode. The display shows “CA”
momentarily and the displayed reading flashes. Use the 

and 

keys to match

the displayed value to the temperature of your solution.

4. 

Press ENTER key.

5. 

Press MODE key to return to pH measurement mode.

Hold function

To freeze your reading, press the HOLD key once.
The LCD show "HO" to indicate the HOLD func-
tion is activated. See figure

Press HOLD key again to deactivate the HOLD
function and return to measurement mode.

pH

A

 C

B

HO

A

A

B

A

5.3

5.4

5.5

Содержание ION 5 Acorn Series

Страница 1: ...ON alone will determine if the product problem is due to deviations or customer misuse Out of warranty products will be repaired on a charge basis 11 Return of items Authorization must be obtained fro...

Страница 2: ...on 12 4 5 Resetting the meter to factory calibration values 13 5 Measurement 14 15 5 1 Measuring ion concentration 14 5 2 Measuring mV 14 5 3 Measuring pH and temperature 15 5 4 Measuring pH with manu...

Страница 3: ...ies Warranty and Return of Items section 2 Display and Keypad Functions The Acorn has a large custom LCD with the following indicators pH HO LO C mV indicates operation mode pH annunciator ION 6 only...

Страница 4: ...tor push together and twist the electrode connector 1 2 turn until it clicks into place To remove simply rotate the connector counter clockwise until it unlocks and slide the connector off the socket...

Страница 5: ...libration This instrument is capable of up to three point ion calibration Calibration values are 0 1 1 0 10 0 or 100 0 You must calibrate to at least two points All calibration points should be one de...

Страница 6: ...s in the solution can affect the calibration and the accuracy of the measurements Before use remove the electrode soaker bottle If the electrode has been stored dry condition the glass bulb by soaking...

Страница 7: ...he and keys to adjust the temperature shown to match the accurate temperature maximum adjustment is 5 C from factory default 6 Press ENTER key to confirm calibration The LCD displays CO and the meter...

Страница 8: ...in pH and temperature 1 Rinse both electrode and temperature sensor with distilled water 2 Turn meter on ION 6 meter automatically enters ion measurement mode Press MODE key until meter enters pH meas...

Страница 9: ...ed solution for about 10 minutes Allow the electrode to cool in some unheated KCl solution Protein deposits prepare a 1 pepsin solution in 0 1M of HCl Set the pH electrode in the solution for five to...

Страница 10: ...ps and dings Also features meter stand for convenient tabletop use Shpg wt 0 5 lb 230 g Ion selective electrodes and solution kits WD 35802 00 Ammonia ion selective electrode includes 15 ml of referen...

Страница 11: ...alibration solution pouches 20 box 0 01 pH accuracy at 25 C Shpg wt 1 1 lb 454 g per box WD 35653 00 Deionized rinse water solution pouches WD 35653 01 pH 4 01 Singles buffer solution pouches WD 35653...

Страница 12: ...23 22...

Отзывы: