Oakton ION 5 Acorn Series Скачать руководство пользователя страница 11

21

20

WD-35812-42 Sodium ion selective electrode, 

includes 15 ml of reference electrolyte

and a filling pipette. Shpg wt 0.5 lb (230 g).

WD-35812-93 Sodium ion selective solution kit,

includes 475 ml Ion Strength

Adjuster, 475 ml calibration standard, 60 ml reference electrolyte, and syringe
dispenser. Shpg wt. 1 lb (460 g). 

WD-35812-48 Water hardness ion selective electrode, 

includes 15 ml of reference

electrolyte and a filling pipette. Shpg wt 0.5 lb (230 g).

WD-35812-99 Water hardness ion selective solution kit,

includes 475 ml Ion

Strength Adjuster, 475 ml calibration standard, 60 ml reference electrolyte, 
and syringe dispenser. Shpg wt. 1 lb (460 g). 

pH electrodes and accessories

WD-35801-00 Replacement electrode,

5.75"L x 0.47 OD (12 mm). 

Shpg wt. 0.5 lb (230 g).

WD-35805-05 Double junction electrode,

use with solutions that are dirty, have

heavy metal or organic ions. 5.75"L x 0.47 OD (12 mm). Shpg wt. 0.5 lb (230 g).

WD-35613-05 Replacement temperature probe, 

316 SS, polypropylene cap, 3" cable.

Shpg wt 0.15 lb (70g).

WD-35805-13 ORP electrode,

epoxy body, single junction, 5.75"L x 0.49"OD 

(12.5 mm). Shpg wt 0.5 lb (230 g).

WD-00653-04 Electrode storage solution,

1 pint bottle. Keeps electrode bulb moist

for faster, more accurate readings. Shpg wt 1.1 lbs (510 g).

WD-00653-06 Electrode cleaning solution,

1 pint bottle. Removes buildup from 

electrodes and maintains electrode sensitivity. Shpg wt 1.1 lbs/510 g.

OAKTON pH calibration solutions

±0.01 pH accuracy at 25°C.  Shpg wt 1.1 lb (510 g).

WD-00654-00 pH 4.01 calibration buffer, 

1 pint.

WD-00654-04 pH 7.01 calibration buffer, 

1 pint.

WD-00654-08 pH 10.01 calibration buffer,

1 pint.

OAKTON “Singles”  pH calibration solution pouches  

20/box.

±0.01 pH accuracy at 25°C. Shpg wt 1.1 lb (454 g) per box.

WD-35653-00 Deionized rinse water solution pouches, 

WD-35653-01 pH 4.01 “Singles” buffer solution pouches

WD-35653-02 pH 7.00 “Singles” buffer solution pouches

WD-35653-03 pH 10.00 “Singles” buffer solution pouches

WD-35653-04 Assortment pack, 

5 ea. deionized water, pH 4.01, pH 7.00, 

and pH 10.00 solution pouches.

NOTE:  Remember to check the temperature calibration when replacing the ATC
probe.  See section 4.4, "Temperature calibration" (page 12).

Appendix A: temperature curve selection (Ion 6 meter only)

Using the temperature calibration procedure described on page 12 of this manual,
you can achieve the ±0.5°C accuracy specified for this meter. However, to 
optimize temperature performance, you may make an offset adjustment of the 
temperature curve.

1. 

Connect your temperature probe to the meter.

2. 

Go to the pH calibration mode by pressing the
CAL key when the unit is in pH measurement
mode.

3. 

Compare displayed value to a NIST certified
thermometer or other thermometer known to
be accurate. For best accuracy, place probe and
thermometer in a constant temperature bath, in
this example: 25°C.

4. 

Press MODE key. Unit displays the default
curve temperature, in this example: 25.9°C

See figure

5. 

Press the MODE key to scroll through the three
temperature curves. Find the temperature
reading closest to your actual temperature, in
this example: 24.8°C.

See figure

6. 

Press ENTER key to confirm calibration.
The LCD flashes "CO",  the meter then reverts 
to temperature calibration mode for offset
adjustment.

See figures

8. 

Use the 

and 

keys to fine tune the 

temperature reading to your actual 
temperature, in this example: 25°C.

6. 

Press ENTER key to confirm calibration.
The LCD displays "CO", and the meter 
then reverts to measurement mode.

 C

A

A

 C

B

B

 C

C

C

D

D

Содержание ION 5 Acorn Series

Страница 1: ...ON alone will determine if the product problem is due to deviations or customer misuse Out of warranty products will be repaired on a charge basis 11 Return of items Authorization must be obtained fro...

Страница 2: ...on 12 4 5 Resetting the meter to factory calibration values 13 5 Measurement 14 15 5 1 Measuring ion concentration 14 5 2 Measuring mV 14 5 3 Measuring pH and temperature 15 5 4 Measuring pH with manu...

Страница 3: ...ies Warranty and Return of Items section 2 Display and Keypad Functions The Acorn has a large custom LCD with the following indicators pH HO LO C mV indicates operation mode pH annunciator ION 6 only...

Страница 4: ...tor push together and twist the electrode connector 1 2 turn until it clicks into place To remove simply rotate the connector counter clockwise until it unlocks and slide the connector off the socket...

Страница 5: ...libration This instrument is capable of up to three point ion calibration Calibration values are 0 1 1 0 10 0 or 100 0 You must calibrate to at least two points All calibration points should be one de...

Страница 6: ...s in the solution can affect the calibration and the accuracy of the measurements Before use remove the electrode soaker bottle If the electrode has been stored dry condition the glass bulb by soaking...

Страница 7: ...he and keys to adjust the temperature shown to match the accurate temperature maximum adjustment is 5 C from factory default 6 Press ENTER key to confirm calibration The LCD displays CO and the meter...

Страница 8: ...in pH and temperature 1 Rinse both electrode and temperature sensor with distilled water 2 Turn meter on ION 6 meter automatically enters ion measurement mode Press MODE key until meter enters pH meas...

Страница 9: ...ed solution for about 10 minutes Allow the electrode to cool in some unheated KCl solution Protein deposits prepare a 1 pepsin solution in 0 1M of HCl Set the pH electrode in the solution for five to...

Страница 10: ...ps and dings Also features meter stand for convenient tabletop use Shpg wt 0 5 lb 230 g Ion selective electrodes and solution kits WD 35802 00 Ammonia ion selective electrode includes 15 ml of referen...

Страница 11: ...alibration solution pouches 20 box 0 01 pH accuracy at 25 C Shpg wt 1 1 lb 454 g per box WD 35653 00 Deionized rinse water solution pouches WD 35653 01 pH 4 01 Singles buffer solution pouches WD 35653...

Страница 12: ...23 22...